Chapter 4 Signal Connections
National Instruments Corporation
4-9
AT-MIO/AI E Series User Manual
DIO<0..7>
DIO
—
Vcc +0.5
13 at (Vcc -0.4)
24 at
0.4
1.1
50 k
Ω
pu
1
SCANCLK
DO
—
—
3.5 at (Vcc -0.4) 5 at 0.4
1.5
50 k
Ω
pu
EXTSTROBE*
DO
—
—
3.5 at (Vcc -0.4) 5 at 0.4
1.5
50 k
Ω
pu
PFI0/TRIG1
ADIO
10 k
Ω
Vcc +0.5/
±
35
3.5 at (Vcc -0.4) 5 at 0.4
1.5
50 k
Ω
pu
2
PFI1/TRIG2
DIO
—
Vcc +0.5
3.5 at (Vcc -0.4) 5 at 0.4
1.5
50 k
Ω
pu
PFI2/CONVERT*
DIO
—
Vcc
+0.5
3.5 at (Vcc -0.4) 5 at 0.4
1.5
50 k
Ω
pu
PFI3/GPCTR1_SOURCE
DIO
—
Vcc +0.5
3.5 at (Vcc -0.4) 5 at 0.4
1.5
50 k
Ω
pu
PFI4/GPCTR1_GATE
DIO
—
Vcc +0.5
3.5 at (Vcc -0.4) 5 at 0.4
1.5
50 k
Ω
pu
GPCTR1_OUT
DO
—
—
3.5 at (Vcc -0.4) 5 at 0.4
1.5
50 k
Ω
pu
PFI5/UPDATE*
DIO
—
Vcc +0.5
3.5 at (Vcc -0.4) 5 at 0.4
1.5
50 k
Ω
pu
PFI6/WFTRIG
DIO
—
Vcc +0.5
3.5 at (Vcc -0.4) 5 at 0.4
1.5
50 k
Ω
pu
PFI7/STARTSCAN
DIO
—
Vcc +0.5
3.5 at (Vcc -0.4) 5 at 0.4
1.5
50 k
Ω
pu
PFI8/GPCTR0_SOURCE
DIO
—
Vcc +0.5
3.5 at (Vcc -0.4) 5 at 0.4
1.5
50 k
Ω
pu
PFI9/GPCTR0_GATE
DIO
—
Vcc +0.5
3.5 at (Vcc -0.4) 5 at 0.4
1.5
50 k
Ω
pu
GPCTR0_OUT
DO
—
—
3.5 at (Vcc -0.4) 5 at 0.4
1.5
50 k
Ω
pu
FREQ_OUT
DO
—
—
3.5 at (Vcc-0.4)
5 at 0.4
1.5
50 k
Ω
pu
1
DIO <6..7> are also pulled down with a 50 k
Ω
resistor.
AI = Analog Input
DIO = Digital Input/Output
pu = pullup
AO = Analog Output
DO = Digital Output
ADIO = Analog/Digital Input/Output
2
Also pulled down with a 10 k
Ω
resistor.
Note:
The tolerance on the 50 k
Ω
pullup and pulldown resistors is very large. Actual value may range between
17 k
Ω
and 100 k
Ω
.
Table 4-1.
I/O Signal Summary, AT-MIO-16E-1, AT-MIO-16E-2, and AT-MIO-64E-3 (Continued)
Signal Name
Drive
Impedance
Input/
Output
Protection
(Volts)
On/Off
Source
(mA at V)
Sink
(mA at
V)
Rise
Time
(ns)
Bias