locations only. Follow these guidelines if you are installing the
NI 9775 in a potentially explosive environment. Not following
these guidelines may result in serious injury or death.
Caution
Do not disconnect I/O-side wires or
connectors unless power has been switched off or the
area is known to be nonhazardous.
Caution
Do not remove modules unless power has
been switched off or the area is known to be
nonhazardous.
Caution
Substitution of components may impair
suitability for Class I, Division 2.
Caution
For Division 2 and Zone 2 applications,
install the system in an enclosure rated to at least IP54
as defined by IEC/EN 60079-15.
Special Conditions for Hazardous Locations Use in
Europe and Internationally
The NI 9775 has been evaluated as Ex nA IIC T4 Gc equipment
under DEMKO 12 ATEX 1202658X and is IECEx UL 14.0089X
certified. Each NI 9775 is marked II 3G and is suitable for use
in Zone 2 hazardous locations, in ambient temperatures of -40 °C