background image

© 2010–2011 National Instruments Corporation. All rights reserved. 

375530B

Aug11

LabVIEW, National Instruments, NI

ni.com, the National Instruments corporate logo, and the Eagle logo are trademarks of National Instruments 

Corporation. Refer to the 

Trademark Information 

at 

ni.com/trademarks

 for other National Instruments trademarks. Other product and 

company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments 
products/technology, refer to the appropriate location: 

Help»Patents

 in your software, the 

patents.txt

 file on your media, or the 

National 

Instruments Patents Notice

 at 

ni.com/patents

. Refer to the 

Export Compliance Information 

at 

ni.com/legal/export-compliance

 

for the National Instruments global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data.

If the latches do not engage, ensure that the expansion bridge is inserted in the orientation 
shown in Figure 2. The expansion bridge fully inserts only with the correct edge up.

Note

The NI 2806 expansion bridge should not be used above the operational shock 

limit. Refer to the NI SwitchBlock carrier or card specifications for these limits. 
Remove or secure the NI 2806 prior to shipping and prior to use in a high-shock 
environment.

Note

To prevent unintentional shorts of source channels between hardwire analog 

bus paths, you cannot use any NI SwitchBlock carrier connected by the NI 2806 
expansion bridge until all connected carriers have been loaded by the driver. Failure 
to load the driver on any carrier connected by the expansion bridge will disable all 
other carriers connected to the same bus. 

Removing the NI 2806 Expansion Bridge

Complete the following steps to remove the NI 2806 expansion bridge:

1.

Power off the chassis.

2.

Squeeze the expansion bridge latches and pull gently on the expansion bridge.

3.

Remove the carriers that were connected to the expansion bridge.

4.

Reinstall the expansion bridge connector covers with the provided 2-56 × 3/16 inch long, 
pan head screws.