Supplies and Replacement Parts
5
Disinfecting the Airway
To disinfect, repeat the standard cleaning procedure, but this
time use a bleach solution instead of soapy water, as specified
by the Centers for Disease Control. Fill the system with the
solution until it reaches the corners of the mouth.
Remember to start filling with the head flat and finish with the
neck slightly elevated to ensure that the solution completely
fills all airway passages. Once completely filled with the bleach
solution, allow the head to sit for at least 10 minutes. Capping
the airway with the supplied red caps will hold the fluid in the
simulator. Drain as described earlier and repeat the rinsing
process to flush out all the bleach solution. Set the head aside
and allow it to dry completely.
Cleaning the Face
Normal soil can be removed from the face of the trainer with
mild soapy water. Nasco Cleaner (LF09919) will remove
stubborn stains. Simply apply Nasco Cleaner to the soiled area
and wipe clean with a soft cloth or paper towel.
Note: Do not use Nasco Cleaner around the mouth and nostrils,
as the residue of the cleaner could be toxic.
Caution: Never place the trainer on any kind of printed paper or
plastic. These materials will transfer indelible stains. Ballpoint
pens will also make indelible stains. Do not use any cosmetics.
Important — Storage of Your Simulator:
The unit must be stored as you received it to avoid damaging
the simulator.