70/ 99
3.
Prepare the patient and connect patient cable to
patient as shown in ‘Fig 9.1 Electrodes placement’.
4.
Select Recording (REC) mode by using the slide
switch.
5.
The green (Recording status) LED on front will glow.
Buzzer will give sound.
6.
Press “Start key”. Buzzer will become off. Memory Full
(Red) LED will glow, and previously stored data will
start erasing. At that time Green LED will remain glow.
7.
After about 5 to10 sec, buzzer will give beep sound
and recording will start. This will be indicated by
blinking of Recording status (Green) LED and Memory
Full (Red) LED will become off.
8.
After 24 hrs. Of recording, Buzzer will give two beeps.
Recording status (Green) LED will stop blinking.
Memory Full (Red) LED will glow continuously
indicating recording is over.
9.
Keep the slide switch at OFF position when data
recording is complete.
10.
When any one of the lead from patient body get
displaced the LEADS OFF LED will glow with buzzer
sound, check lead cable for any loose connection and
press Event Mark key on the unit.
Steps for Data downloading:
1.
Remove batteries from the unit.
2.
Connect download cable to PC and Holter unit.
Summary of Contents for Reveala3 XT
Page 8: ......
Page 57: ...57 106 Fig 7 2 Summary report Part 1 ...
Page 68: ...68 106 Fig 7 16 Poincare Graph and Normal RR histogram ...
Page 73: ...73 99 Reveala 3 with keyboard unit photo USB Data Cable Patient Cable ...
Page 80: ...80 106 Click Next Click Yes Setup ...
Page 81: ...81 106 Click Next Click Next ...
Page 84: ...84 106 ...