7  Specifications 

34 

Nardalert S3 

7.2  Sensor specifications 

Sensor Module 

2271/11 

IEEE 

2271/01 

FCC 

2271/22 

SC6 

2271/31 

ICNIRP 

Frequency Range 

3 MHz to 

50 GHz 

100 kHz to 50 GHz 

Field Measured 

Electric Field, V

2

/m

2

 

Sensor Design 

Radial Field, Diode-Dipole and Thermocouple Array 

Alarm Accuracy 

+4.5 / -3.0 dB (100 kHz to 30 GHz) 

+2.5 / -6.0 dB (30 to 50 GHz) 

+2.5 / -6.0 dB (50 to 100 GHz, Typical) 

Monitor Range 

a

 

5% to 200% of Standard or Guidance 

Immunity at 50/60 Hz 

100 kV/m 

CW Overload 

3000% of Standard or Guidance 

Peak Overload 

32 dB above Standard or Guidance 

Recommended 

Calibration Interval 

2 Years for Sensors (P/N 2271/XX) 

Notes:

 

a

  Percentages related to the highest power density allowed by Standard or Guidance (Controlled, Occupational). 

 

7.3  Outline drawing 

 

 

 

 

Summary of Contents for nardalert S3

Page 1: ...nardalert S3 Personal Monitor Operating Manual...

Page 2: ...l understood how it operates and consulted with your company s safety officer Narda Safety Test Solutions GmbH Sandwiesenstrasse 7 72793 Pfullingen Germany Names and logo are registered trademarks of...

Page 3: ...ent errors 10 Sensor is not installed or operating properly 11 2 4 Dangers from AC mains charger 11 3 Preparing the Nardalert S3 for use 12 3 1 Unpacking 12 Packaging 12 Items included 13 Transport da...

Page 4: ...Cleaning maintenance repair 28 6 1 Cleaning 28 6 2 Performing calibration 29 6 3 Repair 29 6 4 Verification overview 29 Simplified block diagram 29 Theory of operation 30 Testing RF microwave sensors...

Page 5: ...oitation of the electromagnetic spectrum We also have various engineering considerations as well as regulatory requirements to use the electromagnetic spectrum wisely Various authorities have long def...

Page 6: ...forms measurements for human safety purposes particularly in workplace environments where high electric or magnetic field strengths are likely It can also be configured to function as a stand alone ar...

Page 7: ...gnal words CAUTION WARNING and DANGER Follow all the instructions in order to avoid injuries or death NOTICE Indicates a danger that results in damage to or destruction of the instrument CAUTION Indic...

Page 8: ...t on 1 2 3 Sequence of instructions Indicates a sequence of instructions that must be carried out in the order given Result Indicates the result of carrying out an instruction e g The instrument start...

Page 9: ...is Operating Manual Carefully read this entire Operating Manual before you start using the instrument Keep this Operating Manual so that it is available to everyone who uses the instrument and ensure...

Page 10: ...lectromagnetic fields Very strong electromagnetic fields occur near many radiation sources Observe safety barriers and markings In particular persons with electronic implants must keep away from dange...

Page 11: ...n upscale indication should be noticed on the display when the radio is transmitting close to the sensor housing Before beginning any RF radiation measurement always inform yourself of the frequencies...

Page 12: ...to use the Nardalert S3 3 1 Unpacking 3 2 Battery Installation 3 3 Instrument overview 3 4 Connecting the sensor 3 1 Unpacking Packaging The packaging is designed to be re used as long as it has not...

Page 13: ...Phillips 0 2270 90 06 User s Guide and CD ROM with Software NS3 TS 2270 90 07 Car Charger Adapter USB 5V 2259 92 20 Battery rechargeable 2259 92 25 Cable USB2 0 Master Slave A B mini 0 9m 2260 90 58...

Page 14: ...brought into a warm room It may be damaged if used Wait until all visible condensation has evaporated from the instrument surface to avoid damaging the instrument Note The instrument is not ready for...

Page 15: ...r use Nardalert S3 15 3 3 Instrument overview Top Panel Front Panel Rear Panel Power On Off and Enter Key Down Arrow Decrease Up Arrow Increase Alarm LEDs Interchangeable Sensor Alarm LEDs Calibration...

Page 16: ...rt S3 for use 16 Nardalert S3 Side Panels Mini USB port for charging and data transfer Fiber Optic port for data transfer only Rechargeable and replaceable battery located under door Silicon sleeve fo...

Page 17: ...If in doubt cycle unit off and then on again to perform connection test If the sensor is defective not installed or failing pre test it will cause the system to not proceed to measurement mode Chargi...

Page 18: ...nsors under Ordering Information on page 38 as well as in the data sheet of the NS3 These documents can also be downloaded from the Narda website on the internet at www narda sts com Wrong handling of...

Page 19: ...he NS3 is switched on by depressing and holding the On Off button on the left side of the display The alarm LEDs will illuminate and the vibrator will activate before the splash screen Figure 1 is dis...

Page 20: ...ges not actual screen Normal operation The operation screen Figure 5 displays the detected fields from the low and high frequency sensors separately to provide information to the wearer about the type...

Page 21: ...ng properly Note Do not use this function test for verifying calibration This test is only suitable for checking sensors and their connection to the NS3 The readings displayed on the monitor depend on...

Page 22: ...turn to the Menu Selection Screen without implementing the change 4 If the arrow icon is pointing to the Save and Exit line implement the parameter change save the new parameter in memory and return t...

Page 23: ...larm 1 can be set from 10 to 100 and OFF Alarm 2 can be set from 20 to 200 Factory default is Alarm 1 at 50 and Alarm 2 at 200 Backlight Allows setting of display backlight times to OFF 5 seconds 10 s...

Page 24: ...gain to select item to be modified Item is highlighted to confirm selection 3 Tap either the or arrow to cycle through choices 4 Tap the again to select choice Choice is not confirmed until you Save a...

Page 25: ...ody Some customers prefer to wear the unit mounted to a waist belt while others prefer a lanyard attachment Either method of attachment is acceptable as long as the unit is normally kept close to the...

Page 26: ...onal or Controlled limits Standards and guidance s normally recommend conservative exposure limits that are both time and spatially averaged and relate to a 100 reading on the NS3 However the NS3 cann...

Page 27: ...e the monitor to false alarm 5 4 Using the Nardalert S3 as an area monitor The NS3 functions as an effective stand alone area monitoring device It may be operated continuously from the USB supply whil...

Page 28: ...n damage the Nardalert S3 or the charger Water can damage the charger Do not use solvents to clean the Nardalert S3 and charger As cleaning fluid we recommend the use of lukewarm water to which a drop...

Page 29: ...sales partner You can find the Narda sales partner responsible for you on the Narda website at www narda sts com All NS3 mainframes need to be returned to their local calibration facility at least onc...

Page 30: ...d support The NS3 s design utilizes RF shielding and absorbers to isolate it from reflections or scattering produced by the monitor or the human body In general the monitor cannot detect microwave fie...

Page 31: ...to Category 9 monitoring and control instruments As the manufacturer you can return the device to Narda free of charge for proper environment friendly disposal For more information please contact your...

Page 32: ...32 Nardalert S3 7 Specifications 7 1 Monitor specifications 7 2 Sensor specifications 7 3 Outline drawing 7 4 Declaration of conformity 7 5 Declaration of origin...

Page 33: ...erial Full Duplex 57600 baud USB 2 0 mini B jack Serial Full Duplex 57600 baud connector type RP 02 Recommended Calibration Interval 4 Years for Mainframe P N 2270 01 Battery Type Approximate Life RCR...

Page 34: ...ray Alarm Accuracy 4 5 3 0 dB 100 kHz to 30 GHz 2 5 6 0 dB 30 to 50 GHz 2 5 6 0 dB 50 to 100 GHz Typical Monitor Range a 5 to 200 of Standard or Guidance Immunity at 50 60 Hz 100 kV m CW Overload 3000...

Page 35: ...7 Specifications Nardalert S3 35 7 4 Declaration of conformity...

Page 36: ...7 Specifications 36 Nardalert S3 7 5 Declaration of origin Country of origin Germany...

Page 37: ...7 Specifications Nardalert S3 37...

Page 38: ...S3 8 Ordering information This chapter contains the information needed for ordering the NS3 together with its sensors and accessories 8 1 Nardalert S3 part numbers 8 2 Sensor part numbers 8 3 Optional...

Page 39: ...SENSOR MODULE with FCC Sensor Module 2271 101 with IEEE Sensor Module 2271 111 with SC6 Sensor Module 2271 122 with ICNIRP Sensor Module 2271 131 Option Key Nardalert S3 2270 90 01 8 2 Sensor part num...

Page 40: ...435 Moreland Road Hauppauge NY11788 USA Phone 1 631 231 1700 NardaSTS L3Harris com Narda Safety Test Solutions S r l Via Rimini 22 20142 Milano Italy Phone 39 0258188 1 nardait support L3Harris com Na...

Reviews: