5 Operation and use
ELT-400
5-12
5.3.7
Interpreting the measurement results
Fig. 5-5
Measurement display showing measurement result
No.
Explanation
1
Numerical value of measurement result
The following applies to “Exposure STD” mode:
The percentage displayed refers to the limit value specified in the
selected safety standard. A measured value of “100 %” indicates that
the limit value has been reached, regardless of the frequency and
signal shape.
Caution:
The results are calculated using standardized evaluation
schemes in “Stnd” detection mode only.
2
Measurement value units; depend on the selected operating mode.
3
Operating mode
•
The selected standard (standard’s name, exposure range) is
displayed in “Exposure STD” mode.
•
The end value (RMS value) of the measurement range is
displayed in Field strength mode.
4
Detection mode used
The crest factor (peak/RMS) can be calculated from the RMS and
peak values. This gives the factor
for a pure sine wave signal
(only).