26
Electromagnetic C
ompliance
9
UROSHIELD
®
ELECTROMAGNETIC
IMMUNITY
Table 3
Guidance and Manufacturer’s Declaration –
Electromagnetic Immunity
UroShield
®
is intended for use in the electromagnetic
environment specified below. The customer or user of
the UroShield
®
should assure that it is used in such an
environment.
Immunity test
IEC 60601 level
Compliance
level
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
6 Vrms in ISM bands (6.765
MHz to 6.795 MHz; 13.553
MHz to 13.567 MHz; 26.957
MHz to 27.283 MHz; and
40.66 MHz to 40.70 MHz) and
amateur bands (1.8 MHz to
2.0 MHz, 3.5 MHz to 4.0 MHz,
5.3 MHz to 5.4 MHz, 7 MHz to
7.3 MHz, 10.1 MHz to 10.15
MHz, 14 MHz to 14.2 MHz,
18.07 MHz to 18.17 MHz, 21.0
MHz to 21.4 MHz, 24.89 MHz
to 24.99 MHz, 28.0 MHz to
29.7 MHz and 50.0 MHz to
54.0 MHz)
[V] = 3 Vrms
[V] = 6 Vrms
9.1