44
Technical Data
The specifications given here are typical values of the Nanosurf products.
The exact specifications vary somewhat from instrument to instrument,
and are stored on the installation CD, and can be found in the Calibration
dialog (see Software Reference).
Scan ranges and resolution
Maximum XY-Scan range
typ. 0.5 µm
(1 µm optional)
Maximum Z-range
200 nm
Drive resolution Z
0.003 nm
Drive resolution XY
0.015 nm
gap voltage
±10V in 5mV steps
Set point current
±100nA in 25pA steps
sample size
max. 10mm diameter
The exact values are dependent on the calibration of the piezo elements.
Drive resolution is calculated by dividing the scan range over 16 bits.
easyScan electronics
Max. scan speed:
1800 data points per second
feedback loop bandwidth:
3 kHz
Additional User ADC Input:
Option available
T
ECHNICAL
D
ATA
Summary of Contents for easyScan E-STM