Introduction
General Operation
Sample
preparation
Counting Cell
1. Prepare some raw milk sample, SCC-Solution, tube, pipette, and tips.
2. Thoroughly mix the raw milk sample.
1. Add 100 μL of raw milk sample and 100 μL of SCC-Solution in the tube.
(1:1 ratio)
2. Mix the raw milk sample and SCC-
Solution by turning the tube upside
down 3-5 times.
3. Load the cell sample onto the chip.
Then, wait 1 minute for the sample
settling.
- 2 channel: 23μL
- 4 channel: 13μL
Raw milk
Tube
SCC-Solution
13μL
SOMA Chip 4X
Ensure that no bubbles enter each channel.
!
NOTE
SCC-Solution
100 μL
Milk sample
100 μL