10
Environmental
Requirements
System Installation
Power On and
Initial Display
To ensure correct operation and stable performance, install ADAM-rWBC2 in a location
which meets the following conditions :
•
Use at room temperature.
•
Do not expose the device to direct sunlight.
•
Do not subject the device to direct or continuous vibration.
•
Do not subject the device to intense magnetic or electromagnetic fields.
•
Do not install the device in high-humidity environment.
•
Location of device should be free from corrosive gases or other corrosive substances.
•
Ensure minimal contact with dust or other airborne particles.
•
Allow for 10 cm (4 inches) minimum space around the device for proper airflow.
•
Do not place any objects on the device.
1. Check the connection of ADAM-rWBC2 and power cord.
2. Press the power button for 2~3 seconds.
If you get an error message, please contact your local distributor or [email protected].
If booting is successful and no errors are detected, the home screens will be displayed
as below.
CAUTION
At low temperatures (≤10°C), allow the device to warm up for 10 minutes
at ambient temperature before use.
!
CAUTION
Do not tilt the device too much in the forward when connecting the
power cord.
Do not move the device after connecting power cord.
When you connect the power cord to ADAM-rWBC2 even without power
on the device, it will go through self diagnostic tests.
!