ADAM-rWBC2, residual Leukocyte Counter
19
The total number of leukocytes will be displayed after being automatically calculated by
ADAM-rWBC2 software.
Result Analysis
1.71/μL
Total Number of Leukocytes
General Operation
When 1 cell including dust consists of more than 100 pixels or the number of cells is more
than 160 in 1 frame, this is considered as error frame. If these error frames are more than
5, then the error message [Bad Sample] will be displayed on the screen. Please check if the
sample has any dust or cell clumping and retest.
[Bad Sample]
Error Message
50.58