ADAM-rWBC2, residual Leukocyte Counter
15
1. Leave the ADAM-rWBC Kit at room temperature for 10 minutes before use.
2. Carefully dispense 100 µL of well-mixed RBC or platelet or plasma sample into
a clean test tube.
3. Add 400 µL of reagent into the tube.
4. Mix the tube well
· RBC or Platelet : 100 µL 400 µL reagent
· Plasma: 100 µL 100 µL reagent
5. Load 100 µL mixed sample/reagent onto a r-Slide.
6. Wait 4~7 minutes for sample settling.
7. Insert r-Slide into ADAM instrument and press RUN SAMPLE to start.
8. After approximately 1 minute, the result, calculated as WBC per µL, will be displayed
automatically. It may take about 1 minute longer for auto focus at the initial test.
Sample Preparation
and Testing
General Operation
r-Solution
Sample
100 µL
1.71
/µL
RBC, Platelet 400 µL
Plasma: 100 µL
-Result-