Newtons4th Ltd
www.newtons4th.com
Page 2 of 5
Pressing the SCOPE button, the output
signal can be seen and the scope mode
will have default ‘auto’ trigger. Pressing
the SCOPE button again, scope options
are presented. From here, the trigger
option can be changed to ‘normal’ and
after pressing the home key twice, the
scope screen will return with a cursor
mark (>) on the left axis. Using the
▲ and
▼ keys,
the trigger level can be set to the
peak of modulation and a stable waveform
should be visible. Timebase can be
adjusted using the
◄ and ► keys.
Pressing the GRAPH button, CH1, CH2 or
CH1 and CH2 display can be selected.
While this example has the same signal
on both channels, both channels are
independent and can used to show
different signals.
Now, press the SWEEP button to present
the sweep options.
Set the vales as shown here.
Note: While ‘log’ and ‘linear’ step options
are available, only log steps are presently
implemented