4.7 Parameter List
1) rnt : input temperature maximum range value
-
“PT
Type
Board”
setting range : -99.9
–
299.9 degree (Celsius)
-
“CA
Type
Board”
setting range : -99.9
–
999.9 degree, while using decimal point.
or -99
–
1370 degree, while not using decimal point.
4.8 Setting Range : 00.0
–
99.9 degree (Celsius)
2) ACTP (The value of start temperature timer) :
It's parameter for Start temperature value setting timer.
- Absolute value(Present-fixing temperature) > ActP : Timer start working
- Present temperature > fixing temperature - ACTP value : Timer start working
- Present temperature < fixing tempe ACTP value : Timer start working
* Setting range: 00.0 - 99.9 degree(celsius)
1) Prd (Period) : It is Output period
- Setting range : 1
–
99 sec.
2) P (PROPORTION) : It is fixing value of proportion ratio
- Setting range : 0
–
9999
3) A (ANTI-INTEGRAL) : Anti-Integral setting value
- Setting range : 0
–
6999
4) I (INTEGRAL) : It is fixing value of integral calculus
- Setting range : 0
–
6999
5) D (DIFFERENTIAL) : It is fixing value of differential calculus
- Setting range : 0
–
6999