Page 13 / 15
⚫
Operation
1. Place and install the product.
Install the product at the desired place and check the level in all directions
(side by side, front to back and ground).
2.
Connect the power
Prior to connect the power plug, make sure that the POWER S/W is off
3. Power switch on
Turn on the power switch in line of power cable. Then, digital LED will display current temperature n
chamber after below message passed in short time.
4 Setting temperature
a.
Connect the power, t
hen the LED screen will display the temperature in the chamber.
b.
Press the "TEMPSET" key
, then, the LED screen will flicker and display.
c. Then,
input the desired temperature by pushing UP (▲) and DOWN (▼).
d.
Press "TEMP/SET" key again
after putting the desired value. “SAVE” is shown up on the LED
screen as below.
Note
If you don't press "SET" key after set-up, the new set-up value will not be saved at all.
Set-up Temperature range is
15'C ~ 45'C at ambient 25'C.
5 Calibration for temperature
Please follow up below procedure for calibration in case of discrepancy between actual
value (measured by reliable measurement device) in chamber and displayed value.
Measure Temperature after incubator is stabilized in which takes about more than 2 hours (you
might want to perform this stabilization process at night before home)
Please note that low deviation range such as
±
0.1 ~ 0.3% may not be corrected precisely by
this calibration.