background image

ULTRAPEN

X2

™ PTBT6 Operation Manual

 

Nitrate Concentration and Temperature Pen for Use with your Mobile Device Running the PTBTX2™ App

 

PTBT6OM-X2 01-20

 

7

 

 

IV.   INTERFERENCE 

Other anions in the sample solution may interfere with nitrate measurement, depending on the species and their relative 
levels of nitrate. The impact of interference ions on measurements can be expressed by the Nikolsky Equation: 

 

𝐸 = 𝐶𝑜𝑛𝑠𝑡𝑎𝑛𝑡 +

*.,-,./

0

1

2

log (𝑎

 

+ 𝐾

9

 

𝑎

9

:; 0

1

<

Where: 

𝐸 

- the voltage sensor measures 

 

𝑎

 

- the activity of nitrate 

 

𝑎

9

- the activity of interference ion 

 

𝑧

9

- the charge of the interference ion with sign 

 

𝐾

9

 

–the selectivity coefficient of the interference ion for nitrate measurement 

TABLE 1 

 
 

 

 

Generally,  interference  ions  may  cause  positive  errors  for  the  measurement.  The  higher  selectivity  coefficient  the 
interference  ion  has, the  higher  error  it may  cause  at  a  given  concentration.  The  approximate  selectivity  coefficients  of 
PTBT6 for some anions are listed above. For example, if the activity of Cl- is ten times higher than nitrate in the sample, Cl- 
may cause 7% error for the nitrate measurement, while CH

3

COO

-

 and SO

4

2-

 in the same ratio may only cause 0.7% and 

0.05% error, respectively. 
When testing samples with high interference levels, such as wastewater, chlorinated water, and soil water, the interference 
ions  will  be  counted  as  nitrate  with  the  coefficient  factor  and  cause  measurement  errors.  Some  commonly  existing 
interference ions, such as Cl

-

, Br

 -

, I

-

, S

2-

, HCO

3

-

 and NO

2

-

 can be reduced by using Nitrate Interference Suppression Buffer 

(NISB). Add NISB at 1:1 volume ratio to all sample/standard solutions (e.g. add 5 mL NISB to 5 mL solution) in place of 
Nitrate Ionic Strength Adjuster (NISA) throughout the sensor conditioning, calibration and measurement. 

NOTES: 

  Nitrate Interference Suppression Buffer (NISB) must be  purchased separately (see the ACCESSORIES section below). 

  When using NISBSOL, do NOT add Nitrate Ionic Strength Adjuster (NISA). White precipitate may form in sample solution 

after mixing with NISBSOL, it will not affect the measurement. 

  Strong interference ions such as ClO

4

-

 and SCN

-

 require their concentrations to be up to 2000x lower than nitrate in the 

sample to avoid high errors. 

For best results when measuring solution at the extremes of the specified temperature or nitrate range allow the PTBT6 to equilibrate by 
submerging the sensor in the sample solution for 1 minute prior to taking a measurement. 

 

ALWAYS

 make sure the nitrate sensor is completely submerged in the solution being tested. 

  For multiple samples, regularly perform a calibration check with reference solution between sample measurements. 

 

V.  MEASUREMENT VALIDATION METHODS 

CALIBRATION CHECK: 

1. 

Choose one standard solution within the testing range, and prepare the solution according to the measurement 
procedure. 

2. 

Take a measurement using the readings in the reference solution. 

  The value should be within 10% of the expected value. 

  If not, the measurement result may not be reliable. 

  The source of error should be identified and the measurement should be redone after correction. 

KNOWN ADDITION RECOVERY CHECK: 

This validation method works well for validating low concentration measurements (1-60 ppm-N): 

  Using concentrated standard solutions to increase the nitrate concentration in the sample by a known value, optimally 

doubles the measured value. 

  Comparing the value with the increase in PTBT6 readings. 

o

  For about 10 mL sample, one spike of standard solution with the 100 µL pipette will increase the sample nitrate 

concentration by roughly 1% of the standard solution concentration. 

o

 

Example

 ~1 ppm increase by one spike of 100 ppm standard solution, ~10 ppm increase by one spike of 1000 ppm 

standard solution. 

o

  See Table 2 below for more examples. 

 

 

 

SO

4

2-

 

Cl

-

 

CH

3

COO

-

 

K(NO

3

-

/x) 

0.00016 

0.007 

0.0007 

Summary of Contents for ULTRAPENX2 PTBT6

Page 1: ...PTBT6OM X2 01 20 Operation Manual ULTRAPENX2 PTBT6 Nitrate Temperature Pocket Tester For use with PTBTX2 App and Your iOS or Android Mobile Device ...

Page 2: ...g 3 Step 4 Choose a Ultrapen Mode Setting 4 NITRATE MEASUREMENT 5 I Measurement Setup 5 II PTBT6 HOLD Mode Measurements 5 III PTBT6 LIVE Mode Measurements 6 IV Interference 7 V Measurement Validation Methods 7 CALIBRATION 8 I General Notes 8 II Preparing Reference Solution 9 III Verifying Calibration 9 IV 2 Point Calibration 9 V 1 Point Calibration 11 VI Performing a PTBT6 Sensor Diagnostic Check ...

Page 3: ...hen conditioning sensor See MAINTENANCE Section below for storage instructions When removing or replacing the cap gently wiggle it back and forth while pulling pushing DO NOT twist 7 CAP STOP DO NOT push the protective cap beyond the cap stop as sensor damage may occur II PTBTX2 APP MEASUREMENT SCREEN 1 MEASUREMENT VALUE FIELD Displays the measured value of the solution When the PTBT6 turns OFF di...

Page 4: ...d of the concentration values you will be measuring USING THE PTBT6 STEP 1 DOWNLOAD THE ULTRAPEN PTBTX2 APP On and look for STEP 2 CONNECT THE PTBT6 TO THE APP 1 If the PTBT6 is OFF press and release the PEN BUTTON to turn the PTBT6 ON and place it in Standby mode 2 On your Device s Home Screen tap the BLUE PTBTX2 icon to open the PTBTX2 App 3 Tap the CONNECT button at the bottom of the App displa...

Page 5: ...ys nitrate in ppm mV Displays the raw signal output of the nitrate sensor in millivolt mV To choose which nitrate parameter to display during measurements 1 If the PTBT6 is OFF press and release the PEN BUTTON to turn the PTBT6 ON and place it in Standby mode 2 Tap the PREFERENCES button in the Feature Navigation Bar 3 Tap the line that says Ultrapen Settings 4 Tap the Parameter Select line 5 The ...

Page 6: ...held on the App s display and the App places the PTBT6 into Standby mode The user may end the LIVE mode measurement series any time prior to that by tapping the BLUE button on the App display To select either the LIVE or the HOLD mode 1 If the PTBT6 is OFF press and release the PEN BUTTON to turn the PTBT6 ON and place it in Standby mode 2 Tap the PREFERENCES button in the Feature Navigation Bar 3...

Page 7: ...REPARING A SAMPLE SOLUTION 1 Fill the graduated cylinder to the 10 mL line with the sample solution 2 Add 200 µL of the Nitrate Ionic Strength Adjuster NISA solution to the graduated cylinder using the supplied 100 µL pipette 3 Replace the cap on the graduated cylinder and shake well to mix the prepared test solution II PTBT6 HOLD MODE MEASUREMENTS 1 If the PTBT6 is OFF press and release the PEN B...

Page 8: ...l flash rapidly 6 When the PTBT6 stabilizes it will begin displaying measurements a A flashing BLUE LIVE mode indictor will appear in the upper right corner of the Measurement Value field b A BLUE button will appear that allows the user to manually end LIVE mode measurements at any time 7 Hold the Ultrapen steady in the sample solution while the LIVE mode measurement is continuing DO NOT SWIRL 8 A...

Page 9: ...SB must be purchased separately see the ACCESSORIES section below When using NISBSOL do NOT add Nitrate Ionic Strength Adjuster NISA White precipitate may form in sample solution after mixing with NISBSOL it will not affect the measurement Strong interference ions such as ClO4 and SCN require their concentrations to be up to 2000x lower than nitrate in the sample to avoid high errors For best resu...

Page 10: ...g the PTBT6 before each series of tests depending on usage I GENERAL NOTES Nitrate sensor should be conditioned prior to performing a calibration see SENSOR CONDITIONING above The PTBT6 offers 1 point or 2 point calibration and is calibrated in nitrate as nitrogen NO3 N using 1 10 100 or 1000 ppm reference solutions For greatest accuracy complete a 2 point calibration selecting the closest referen...

Page 11: ...e reference solution that is closest in value to the expected nitrate values you will be measuring 2 If the reading is off more than 10 from the value of the reference solution calibrate the PTBT6 3 If the reading is still off more than 10 from the value of the reference solution perform a sensor check as defined in Section VI below 4 If the problem persists the sensor should be replaced IV 2 POIN...

Page 12: ... or tissue to remove any water drops 15 Dip the pen in the prepared 1000 ppm reference solution Make sure the sensor is completely submerged 16 Tap the CONTINUE Button The PTBT6 Secondary CAL screen will appear The CAL button corresponding to the initial CAL point will be inactive 17 Tap the button corresponding to the second CAL point 1000 ppm in this example 18 The PTBT6 will begin calibrating t...

Page 13: ...remove any water drops 4 Dip the pen in the prepared 1 ppm reference solution Make sure the sensor is completely submerged 5 Swirl it around to remove any bubbles that may have been caught in the sensor 6 Tap the Calibration button in the Feature Navigation Bar 7 The Initial CAL screen will appear 8 Tap the CAL button 9 The PTBT6 Secondary CAL screen will appear 10 Tap the 1 ppm CAL button 11 The ...

Page 14: ...temperatures outside of this range If the sensor gain is outside of these limits Recondition the sensor as described in the PTBT6 Operation Manual PTBT6OM X2 Redo a 2 point calibration using 100 ppm as the initial CAL point and 1000 ppm final CAL point If the problem persists the sensor should be replaced VII FACTORY CALIBRATION RESET The PTBT6 can be returned to factory default calibration settin...

Page 15: ...K to accept the new email format EXPORTING DATA FILES 1 Tap the MEMORY RECALL button in the Feature Navigation Bar 2 Tap the EDIT button in the top right of the screen 3 The Record list will switch to Edit mode 4 If the records you want to select do not appear on the screen swipe UP or DOWN until you find them 4 Select individual records you wish to email by tapping each record 5 The DELETE and CL...

Page 16: ... then carefully pat the sensor with a clean soft lint free cloth to remove any water drops 2 Rinse the protective cap with clean water preferably DI RO or distilled 3 Pat with a clean soft lint free cloth to remove any water drops 4 Put the protective cap on the sensor and store the PTBT6 in a cool place III REPLACING OR REINSTALLING THE NITRATE SENSOR Follow the instructions that come with your r...

Page 17: ...the Ultrapen is ON and in Standby mode 3 Check the status of the mobile device s Bluetooth feature If it is OFF turn it ON If it is ON turn it OFF then back ON 4 Wait for 10 seconds while the mobile device rescans for nearby Bluetooth devices 5 Reopen the PTBTX2 App 6 Tap tHE CONNECT DISCONNECT button 7 Swipe down on the Ultrapen list 8 The Ultrapen should appear on the list The battery of Mobile ...

Page 18: ...as nitrogen in parts per million ppm NO3 Displays nitrate in ppm mV Displays the raw signal output of the nitrate sensor in millivolt mV TABLE 7 ULTRAPENX2 PTBT6 Performance Specifications Nitrate as nitrogen NO3 N Nitrate NO3 Temperature Ranges 1 1400 ppm 4 4 6200 ppm 0 71 C 32 160 F Accuracy1 10 of the reading 10 of the reading 0 1 C Resolution 0 1 ppm 0 1 ppm 0 1 C F pH Range 3 to 11 pH Time to...

Page 19: ...ate as nitrogen Reference Solution N10SOL 10 ppm NO3 N nitrate as nitrogen Reference Solution N100SOL 100 ppm NO3 N nitrate as nitrogen Reference Solution N1000SOL 1000 ppm NO3 N nitrate as Nitrogen Reference Solution IONIC STRENGTH ADJUSTER Order Model NISASOL NITRATE INTERFERENCE SUPPRESSION BUFFER Order Model NISBSOL REPLACEMENT NITRATE SENSOR Order Model RPT6 MYRON L WARRANTY The PTBT6 excludi...

Page 20: ...01 20 MYRON L COMPANY 2450 Impala Drive Carlsbad CA 92010 7226 USA Tel 1 760 438 2021 Fax 1 760 931 9189 E Mail info myronl com techquestions myronl com www myronl com Designed Engineered and Assembled in the USA ...

Reviews: