ULTRAPEN
X2
™ PTBT6 Operation Manual
Nitrate Concentration and Temperature Pen for Use with your Mobile Device Running the PTBTX2™ App
PTBT6OM-X2 01-20
16
TABLE 5 (continued)
Symptom
Possible Cause
Remedy
Measurement readings are not as
expected.
Sensor was not properly
submerged in solution during
measurement.
Repeat measurement and ensure sensor is
submerged in solution prior to LED flashing quickly.
Sensor needs conditioning.
Condition sensor (see MAINTENANCE section).
Membrane is damaged.
Replace the sensor (see MAINTENANCE section).
Sensor is damaged.
If cleaning sensor does not correct error, replace
sensor (see MAINTENANCE section).
Measurement screen displays “Out
of Range”.
Possible sensor damage.
1. Perform a measurement on a known, good
sample of prepared reference solution.
2. If the result is still the “Out of Range” message,
replace the sensor (see MAINTENANCE section).
Error message during calibration.
Sensor was not properly
submerged in solution during
calibration.
Repeat calibration and ensure sensor is submerged
in solution prior to LED flashing quickly.
Improper reference solution.
Verify you are using the proper reference solution.
Temperature out of range.
Verify temperature is within range.
Sensor needs conditioning.
Condition the sensor (see MAINTENANCE section).
Sensor is damaged.
If conditioning the sensor does not correct error,
replace sensor (see MAINTENANCE section).
SPECIFICATIONS
TABLE 6
ULTRAPEN
X2
PTBT6 Measurement Types (parameters)
Parameter
Measurement
NO
3
–
-N
Displays nitrate as nitrogen in parts per million (ppm).
NO
3
Displays nitrate in ppm.
mV
Displays the raw signal output of the nitrate sensor in millivolt (mV)
TABLE 7
ULTRAPEN
X2
PTBT6 Performance Specifications
Nitrate as nitrogen NO
3
-
-N
Nitrate
NO
3
-
Temperature
Ranges
1-1400 ppm
4.4-6200 ppm
0-71˚C / 32-160˚F
Accuracy
1
±10% of the reading
±10% of the reading
± 0.1°C
Resolution
0.1 ppm
0.1 ppm
0.1°C/F
pH Range: 3 to 11 pH
Time to Reading Stabilization: 10 - 180 seconds
Power Consumption: Active Mode: 30 - 140 mA; Standby Mode: 2 µA
Physical Dimensions: 17.15 cm L x 1.59 cm D or 6.75 in. L x .625 in. D
Weight: 54 g or 1.94 oz.
Case Material: Anodized Aircraft Aluminum with Protective Coating
Battery Type: One N type, Alkaline 1.5 V
Calibration reference solutions: 1, 10, 100, or 1000 ppm NO
3
-
-N
Operating / Storage Temperature: 0-40ºC / 32-104ºF
2
Mobile Device System Requirements: The PTBTX2 App requires a mobile device running either iOS 10.0 or later; or Android 7.0 or later.
EPA Compatibility: The PTBT6 is compatible with EPA Approved Standard Method 4500-NO3
-
D
Enclosure Ratings: IP67 and NEMA 6
EN61236-1: 2006 - Annex A: 2008: Electrostatic discharge to the PTBT6 may cause it to spontaneously power ON.
In this case, the PTBT6 will power OFF after several seconds.
WARNING!
These products can expose you to chemicals including Di(2-ethylhexyl)phthalate (DEHP), which is known to the
State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
1
Accuracy of measuring reference solutions after Temperature Controlled 2-Point Calibration with one order of concentration difference.
2
For extended life of the sensor, store the PTBT6 in a cool location (5-25˚C).
DO NOT FREEZE
!