17. Those who have cesarean section within half a year or those who have abortion within 3
months or natural delivery within 2 months can not use this machine.
18. If the suction is set at too high level accidentally during the treatment, you should stop the
treatment immediately and reset the energy level.
19. Use machine or train machine operators in strict accordance with instructions in the user
manual.
4. Troubleshooting & Solutions
1. The instrument cannot be started, and the button light on the back of the instrument does
not work?
A. Make sure the power cord is connected to the socket with normal power supply.
B. Whether the fuse on the back of the instrument is loose or blown.
2. No RF output from the instrument?
A. Please check whether the cord connects the handle to the machine properly
B. Please check whether the treatment area has been cleaned. Grease or essential oil products
may cause poor contact between the handle and human body to cause this problem.
3. Weakened RF output?
A. Please check whether the cord connects the handle to the machine properly
B. Please check whether non-conductive grease and other substances are on the treatment
handle, which can cause poor contact between the handle and skin to weaken the output.
C. Please check whether the skin product used is specified by the instrument.
4. No suction or very little suction?
A. Please turn off the instrument and check whether the filter used in the instrument should be
replaced
B. Please check the rubber seal in the handle, because the rubber seal may be worn to cause air
leakage.
C. Please check whether the filter’s container is screwed tight and whether the rubber seal is
worn, for air leakage may lead to poor air pressure.
D. If the methods above cannot solve this problem, please contact the instrument dealer for
assistance.
5. I can start the instrument, but the screen shows erroneous message?
A. Please unplug the power cord behind the instrument and wait for about 1 minute before
restarting the instrument.
B. If the above method cannot solve the problem, please contact the instrument dealer for
assistance.
5. FAQs
1. Q: How long can I start seeing results of RF treatment?
A: Normally, you can see the treatment results that day when you take the operation or within
one week. Skin collagen tissue is heated and then contracts, and you can obviously feel the
sensation of skin being tightened. Because RF works to stimulate subcutaneous tissue to
promote collagen regeneration, so the more you take the treatment, the more noticeable the
effects will be.
Summary of Contents for MS-54K2