andard Treatment Process

2.1 Form filling

Fill in the blanks of the Client File, confirm the treatment program, remind the client pay

attention to some points before and after treatment.
2.2 Face cleaning

Clean the face with gentle face cream for 3 minutes.
Remind: clean the hands of operator and the handpiece with bromogeramine.

2.3 Preheating

Start the equipment after face cleaning, preheat for not less than 3 minutes (without lighting

the xenon lamp).
2.4 Shaving

Shave the unwanted hair of the treatment area(2 minutes).

2.5 Gel coating

Coating the gel evenly (2-3mm) on the skin.
Effect

light guide, cooling, anaesthetisin.

Attention: no coating on eyes, eyebrows, lips.

2.6 Eye protection

Operator: wear the safety goggles.
Patient: close the eyes, wear professional safety goggles or cover the eyes with eye patch.

2.7 Allergic test

Emit 2 or 3 light spots in the bottom of ears in safe parameter, observe for 3 minutes.
Effect: do not continue to operate if there is a allergic phenomena.

2.8 Operation

Carry on the operation with a suitable parameter to the operatee.

Attention

2.8.1 Inform the operatee that there will be a slight pricking in operation before operating, to
avoid panic.
2.8.2 Start the operation with a low energy grade, adjust the parameters in operating while asking
the feeling of the operatee.

Summary of Contents for HR-T350

Page 1: ...Page 1 in 12 pages Please read the instructions before use Brief Introduction E light applies three core technology IPL RF and epidermis cooling It combines the USER MANUAL...

Page 2: ...hermolysis effect Since light energy is far below conventional IPL the pigment tissue and vascular pathological tissue of the corium and epidermis would absorb more energy without injuring the epiderm...

Page 3: ...ense pulsed light source xenon lamp B Reflector and filter C Sapphire wave guide 2 Technique Specification Light Energy 1 50J cm2 RF energy 1 25J SR Spectra 530 1200nm Cooling water air semi conductor...

Page 4: ...ing the gel evenly 2 3mm on the skin Effect light guide cooling anaesthetisin Attention no coating on eyes eyebrows lips 2 6 Eye protection Operator wear the safety goggles Patient close the eyes wear...

Page 5: ...ction before for similar operation fill the Client File concretely Adjust the parameters on the basis of full acquaintance of the condition of operatee Inform the operatee the possible reation in oper...

Page 6: ...3 4 2 Double confirm there is no leaking of water circulation fold the protective cover by turning it right 3 5 Debugging 3 5 1 Make sure water and electric circuit is correct connected then turn on t...

Page 7: ...e wanted treatment program by clicking 3 6 2 Operation Interface Operation Interface Treatment mode 3 groups of different parameters setting Pulse1 working time of the first pulse Pluses sub pulse wid...

Page 8: ...amp is on but could not make shots start pause click start zimmer the handle and the capacitor charge automatically system will enter the treatment status when charge is ready Click stop escape from t...

Page 9: ...Blackhead and oil comeout in 2 3 days and disappear in 7 days Parameter control Long pulse width short interval high energy 3 8 7 Winkle pouch black circle Light pain Parameter control Short pulse wi...

Page 10: ...t by improper handling Operators should keep vigilant and take necessary protection measures according to the manual Operatee Carry out the treatment in the professional treatment room take effective...

Page 11: ...wer therefore do not dismantle the backboard without under supervising by our technicians Do not move the equipment by pulling the handpiece Do not use the equipment in inflammable and explosive envir...

Page 12: ...tion and user manual The warranty does not extend to any equipment that a Malfunctions caused by abused operation b Malfunctions caused by modification or repair by anyone except EASYLASER or not auth...

Reviews: