COUNTING FUNCTION
1. Press , the display will show “ 10Cnr” (means sample size is 10 pcs)
2. Press again and again, “10
、” “
20
、” “
50
、” “
100” pcs will appear in succession.
Stop at the one you want to use.
3. Put the exact quantity of samples desired on the platform and press , the set sample size will appear.
4. Keep adding objects to be counted on the pan, the total number of the objects will be displayed.
If the unit weight is too small for the counting resolution, the display will show “ErrPcS”
ERROR MESSAGES
When an applied load exceeds the capacity, “
-----9
” will appear on the display. Remove excessive load immediately.
The unit may return to normal operation. Remember: You can permanently damage the scale and void your warranty
by overloading it!
The LCD will display “
ERR 1
” if the weight placed on the platform is incorrect during calibration.
POWER SUPPLY
The iMO1™ is powered by the
DC 6V/1.3Ah rechargeable sealed lead-acid battery
or directly by the
8.5V / 0.2A AC power adaptor
.
When the symbol appears, it indicates that the battery needs to be recharged. The balance will automatically
CALIBRATION
Press and hold (do not release) and power the scale on with , release when the LCD shows “CAL”.
2. Press again, it shows “000000”. This is where you would input the calibration weight you are going to use to
calibrate (500 grams is recommended). Use to cycle through the 6 zeros shown on the screen and use on
the selected digit to adjust selected digit. To calibrate a the recommended 500 grams, the screen must read
“000500”. Once this is set, place the 500 gram calibration weight on the tray.
WEIGHT RESPONSE SPEED
Press and hold (do not release) and power the scale on with . Wait until the display shows “nb0”, “nb1”,
“nb2”, or “nb3”. You can now release . Press again to select your response speed. (nb0: slowest, nb3 fastest)
RANGE OF ZERO TRACK AND ZERO DISPLAY SELECTION
1. Press and hold (do not release) and power the scale on with . Wait until the display shows “0.5d. 1.0d.