*Maintenance:
In case when the water is hard, which means the water contains too much calcium and
magnesium, white scaling can occur on the water basin, transducer and inner surface of the
water tank. The unit will not work effectively if scaling is present. If this occurs, the following
is recommended:
1) Clean the transducer, water tank and basin unit every week.
2) Use cooled boiled or distilled water.
3) Change the water in water tank regularly to keep it fresh.
4) Ensure you remove all water from the water tank and dry all parts including the water basin
before you store the unit unused for a long period of time.
5) Do not use chemicals for cleaning.
How to clean the transducer:
1) Put 5-10 drops of vinegar on the surface of the transducer, wait for 2-5 minutes;
2) Brush out the scale on the surface.
3) Rinse the transducer with clean water.
How to clean the basin unit:
1) Clean the water basin with soft cloth and the vinegar if there is scale in it.
2) If the scale is covered on the security switch, drop the vinegar on the scale and clean it out
with the brush.
3) Rinse the water basin with clean water.
DISPOSAL:
Please dispose of this device responsibly, in according with environmental guidelines of your
country.
Indications: To humidify the indoor environment of the patient in order to get a higher relative
humidity. If the patient shows no improvement, consult a healthcare professional.
Storage Conditions: Store the unused device in a dry place below 70 °C.
Date of issue: 30 November 2019.