MEASUREMENT OF RESISTIVE CURRENT (Ior) 

(At the grounding line) 

1) Set the power switch "ON".
2) By "Ior SELECT" switch, set the circuit condition to Single Phase (1

φ

/Ir) or

Three Phase (

/Ir).

3) Set the range switch to the appropriate range to be measured.

(Select bigger range than the expected current value).

4) Open CT part and clamp it to the grounding line to be measured. Close CT

completely.

5) While reading the displayed value, set the range to the most appropriate one.

(If you can locate the instrument to the direction less influence of external magnetic

field, you can make measurement more accurately). 

6) Press "Ior" switch one time. After approx. 6 seconds, the first Ior value will be

displayed.  (Take  average  value  of  several  times).  By press  "Ior" switch  once  again,
the    mode will return to Io current measurement.

(Except for the grounding line) 

In case of single phase/two or three wires, clamp CT to two wires or three wires 

    together. Setting of circuit is "1

φ

/Ir" by Ior SELECT switch. 

In  case  of  three  phase/three  or  four  wires,  clamp  CT  to  three  or  four  wires  all 
together. 
Setting of circuit is "

/Ir". 

The operation procedure is same as for leakage current of grounding line. 

Note 

Do not mistake to set the circuit condition in case of Ior measurement 

In case of harmonics current and Ior current measurements, cannot detect the values 
unless  the  fundamental  components  (50/60Hz)  are  inputted  more  than  2%  of  full 
scale 
of each range. 

Do not change the range according to the values of harmonics current. It may cause 
errors. 

By Auto Power Off function, the power will get off automatically approx. 10 minutes 
after the final key operation. 

When "                " sign has appeared on the display during the use, it shows the 

    batteries are exhausted. Change the batteries to the new ones immediately. 

Summary of Contents for MCL-400IR

Page 1: ...ad this instruction manual completely and familiarize yourself thoroughly with all functions Keep this instruction manual carefully to take out whenever you need MULTI MEASURING INSTRUMENTS CO LTD Aki...

Page 2: ...or storage this instrument where the mechanical vibration can be directly transmitted as it may cause defect of the instrument Do not use this instrument nearby the appliances which generate strong ma...

Page 3: ...ircumstances there are two kinds of instruments 1 Igr measuring apparatus with low frequency injection method 2 Ior measuring apparatus with simplified effective method In the low frequency injection...

Page 4: ...side of unloading transformer The ratio of the 5th and 7th harmonic voltage at the low voltage side influenced from the high voltage lines is almost fixed Consequently the compositions of the 5th and...

Page 5: ...nge Selection Rotary Switch Input Frequency 45 60Hz AC Conversion RMS Detection Method A D Conversion Double Integration Method Display 3 3 4 digit LCD max reading of 4000 Sampling Rate 2 times second...

Page 6: ...al 1 Batteries 3 installed into the instrument 3 2 ELECTRICAL CHARACTERIC 23 5 less than 80 RH Range Resolution Accuracy AC40mA 0 01mA 1 rdg 8dgt AC400mA 0 1mA AC4A 0 001A AC40A 0 01A AC300A 0 1A 1 rd...

Page 7: ...Switch Manual range switch for the selection of current 40mA 400mA 4A 40A 300A and voltage 500V Data Hold Switch By press one time DH sign will be displayed on LCD and the displayed value will be hel...

Page 8: ...this switch the calculation will start and Wait mark is lightening on the display until the calculation had been finished After that Ior value will be displayed The displayed value before calculation...

Page 9: ...re the line current MEASUREMENT OF LEAKAGE CURRENT 1 Measurement for leakage current of grounding line Same procedure as for the measurement of line current 2 Except for grounding line The operation i...

Page 10: ...current measurement Except for the grounding line In case of single phase two or three wires clamp CT to two wires or three wires together Setting of circuit is 1 Ir by Ior SELECT switch In case of t...

Page 11: ...test leads MEASUREMENT OF HARMONICS VOLTAGE Refer to the operation of MEASUREMENT OF HARMONICS CURRENT But the measuring method is same as MEASUREMENT OF VOLTAGE WARNING It may cause electric shock Ch...

Page 12: ...d It may cause liquid leakage of batteries and may cause the damage to the instrument When the batteries get exhausted and voltage is getting below than the operation Voltage sign will appear on the d...

Page 13: ...contact the dealer where you bought the instrument The warranty period is 12 months from the date of purchase and the instrument shall be repaired at free of charge provided that we judge the cause of...

Page 14: ...lation deterioration of grounding phase In case that the mark is lightening on off during the measurement of resistive leakage current Ior the instrument is detecting the insulation deterioration of g...

Reviews: