SPECIFICATION
PERFORMANCE SPECIFICATION
Current measurement Accuracy
(23.5
℃
, 85% RH or less, no condensation)
At 50Hz / 60Hz (Sine wave) *The conductor is positioned at the center of CT.
Range
Resolution
Measuring Range
Accuracy
30mA
0.01mA
0.10mA~30mA
±
1.2%rdg
±
5dgt
300mA
0.1mA
30mA~300mA
±
1.2%rdg±5dgt
3A
0.001A
0.3A~3A
±
1.2%rdg±5dgt
30A
0.01A
3A~30A
±
1.2%rdg±5dgt
300A
0.1A
30A~300A
(30A~200A) 1.2%rdg±5dgt
(200.1A~250A)
±
3.0%rdg±5dgt
(250.1A~300A)
±
5.0%rdg
±
5dgt
*0 (zero) is displayed for values of 10-counts or less of each range.
*Crest factor: <2
*Temperature and Humidity measurement accuracy; reference value
*Displayed by Multi Tracer (M 1141XR, M 1141X)
GENERAL SPECIFICATIONs
Measurement; AC leakage current (Io), AC load current (I)
Window Diameter;
φ40mm (CT clamp method)
Display; *Liquid crystal display (2 lines, 4 digits, with unit symbol)
*"AUTO": Auto range
*"FL": Filter mode
*"DH": Data hold
*"MAX": MAX hold
*Battery level display
Range; Auto range
Manual range (30mA / 300mA / 3A / 30A / 300A)
Detection Method; (M-1141 / 1141X) RMS-converted mean value rectification
(M-1141XR) True RMS values by the analog calculation