1. OVERVIEW
This instrument is a high-accuracy clamp-type leakage current meter that
greatly improved convenience using the latest technology.
Features
•
Since the influence from the external magnetic field is slight, a minute current
can be measured accurately even near a motor or other wiring.
• The iron core made of a special alloy that does not rust for a long time use
provides less aging stable accuracy.
•
The filter function allows you to know the amount of high frequency content
in the leakage current.
• This instrument has a Bluetooth communication function. It communicates
with a smartphone / tablet with the dedicated application "Multi-Tracer"
installed, and can display measured values or save them on a server. (Only
applicable with MCL-800DX and MCL-800DXR)
2. SPECIFICATIONS
2-1 Performance
2-1-1 Current measurement accuracy
Conditions 23
℃±
5
℃
, 80 %RH (No condensation)
50 / 60 Hz (Sine wave) Location of the wire to be measured: at the center of the CT
Coupling type
: AC coupling
Crest factor
: 2.5 or less up to 200A range (at full scale input)
(Only MCL-800DXR)
1.41 or less in the range of 1000A (at full scale input)
Range
Resolution
Measuring range
Accuracy
20 mA
0.01 mA
0.10 mA to 22.00 mA
±
2 %rdg
±
5 dgt
200 mA
0.1 mA
1.0 mA to 220.0 mA
2 A
0.001 A
0.01 A to 2.200 A
20 A
0.01 A
0.10 A to 22.00 A
200 A
0.1 A
1.0 A to 220.0 A
1000 A
1 A
10 A to 1000 A
* When the value is less than 10 counts in each range, "0" is displayed.