Beluga User manual
1. Attach the T-piece circuit, gas hose
Setup
10
Setup
11
A. Attach the T-piece circuit to the mixed gas outlet in the front of the device.
B. Attach the compressed oxygen source to the CGA 1240 oxygen inlet.
2. Plug in and power up
A. The connector plugs into the side of the device and then directly into the power
outlet in the wall.
B. Press the Power Button to turn the device ON.
Only use T-piece circuit and power supply unit provided by MTTS.
The gas sources must provide pressure within the range of 10 to 60 psi
(0.7 to 4.1 bar). If the gas sources are outside of this range, the device may
not be able to produce the desired %FiO2.
Attaching the compressed oxygen will not waste any oxygen. The machine will
not use any compressed oxygen until you increase the FiO2 setpoint.