![mtts Beluga-V1 User Manual Download Page 11](http://html1.mh-extra.com/html/mtts/beluga-v1/beluga-v1_user-manual_1852604011.webp)
V-0-03-EN
Beluga User manual
Maintenance
Only qualifi ed personnel should carry out service and maintenance proce-
dures.
After the maintenance is completed, ensure the equipment is functioning cor-
rectly in accordance with the published performance specifi cations.
Ensure only approved replacement parts are used during service and mainte-
nance procedures.
Please contact an authorised MTTS representative for further assistance with
any servicing or maintenance requirement.
20
The Beluga Ventilator includes a number of parts that must be replaced or serviced
during the lifetime of the device. These parts include:
Component
Approx. Duration of Use
Maintenance Information
Air fi lter
6 months
Clean or replace.
Oxygen sensor
18 months
Replace if it is fauty or FiO2 control
is not accurate after calibration.
Maintenance
Device information screen
To better understand the device performance, device information should be checked
before opening the device. From Setting screen, touch the Info button to see this in-
formation.
Air fi lter and oxygen sensor replacement
To access air fi lter and oxygen sensor, open the back hatch with a Philips head
screwdriver. Gently take the fi lter/sensor out and replace with the new one.
Air fi lter
Oxygen sensor