background image

 

 

 

Flow: Water flow detection 

 

Temperature

Temperature detection 

 

Working and pause selection button             

 

 

READY: Ready to work 

 

STANDBY

Pause to work 

 

Appears when the READY button is pressed     

         

 

 

Appears when the STANDBY button is pressed            

 

When pressing the READY button, press the handle switch to start 

shot 

 

Alarm display

 

Display the following content in the prompt box

       

 

Temperature: 

Turn red and flashing (alarm when the water 

temperature is higher than 30 degrees) 

 

Flow

  :Turn Red and flashing (alarm when the water flow is 

below 2) 

5. Setting Interface 

  Press System setting

Enter the setting interface 

Ⅴ、

Treatment considerations 

Summary of Contents for D808 PRO

Page 1: ...User manual 808nm Diode Laser Hair Removal Machine D808 PRO...

Page 2: ...CONTENTS 1 Introduction 2 Security 3 System Installation 4 Machine operation 5 Treatment considerations 6 Maintenance 7 Technical description 8 Environmental protection...

Page 3: ...hs warranty The machine is guaranteed within 36 months from the date of shipment Among them the circuit board touch screen and related accessories are covered by the warranty but consumable items such...

Page 4: ...on the skin surface The wavelength of the semiconductor laser can penetrate into the hair follicles and does not damage the skin surface like ruby and alexandrine This allows semiconductor lasers to b...

Page 5: ...Security Motor classification Protection against electric shock Type I Type B Protection level against liquid ordinary equipment Operating mode intermittent and continuous Not suitable for environment...

Page 6: ...operation Optical security This system is the first type of laser product and has a mark on the laser radiation port of the hand tool When operating the equipment the operator should wear protective g...

Page 7: ...ll electricity on some parts Therefore please cover all housings Except the manufacturer and its authorized maintenance personnel no one should disassemble the casing otherwise it may cause electric s...

Page 8: ...orbing substance Do not use the device in an environment with flammable substances Keep the surface of the device clean and use a suitable detergent to clean it before operation If alcohol is used for...

Page 9: ...A Emergency Switch E Water injection hole B Key Switch F Power Switch C Handle G Power line interface D Handle hanger H Overflow hole I Drain hole...

Page 10: ...w the overflow nut before adding water pull out the funnel after filling and do not cover the filling nut Keep the air flow of the water tank smooth Plug the power cord into the corresponding power so...

Page 11: ...se a cart to move the device When moving never tilt the device When moving the device do not push or pull hard you should try to lift the device Machine operation Turn on Turn off Machine switch 1 Bef...

Page 12: ...Connect the power line Turn on the power switch Turn on the emergency stop switch Use the key to turn on the instrument Operation interface Selection Skin selection interface six kinds of skin select...

Page 13: ...Extension button hair color selection...

Page 14: ...working interface Cooling level Cooling regulation 1 4...

Page 15: ...e width range is 10ms 250ms 3Hz Pulse width range is 10 160ms 4Hz Pulse width range is 10 120ms 5Hz Pulse width range is 10 100ms 6Hz Pulse width range is 10 80ms 7Hz Pulse width range is 10 70ms 8Hz...

Page 16: ...Y button is pressed When pressing the READY button press the handle switch to start shot Alarm display Display the following content in the prompt box Temperature Turn red and flashing alarm when the...

Page 17: ...tment 1 Establish customer information files 2 Exclude patients with the above contraindications 3 Determine why the patient seeks treatment and its expectations and tell the patient the course of tre...

Page 18: ...d from a lower setting parameter and energy is gradually increased to avoid side effects When treating light photoaging skin the parameter selection and scar treatment are similar you can refer to the...

Page 19: ...5 and hydrocortisone ointment 1 For patients with skin type III or IV pigmentation may appear within two to four weeks after treatment but it can disappear within three months Daily maintenance Machi...

Page 20: ...umidity ordinary equipment Avoid contact with combustible substances Operation mode continuous First class laser products Environmental protection We have taken environmental requirements into conside...

Page 21: ...ereby removing the hair for a long time Machine advantages 1 Safe without sinking laser pulse light directly hits the hair follicle melanin the epidermis absorbs less energy so there is no loss of sur...

Page 22: ...from various parts of the human body such as armpit hair beard lip hair hair bun bikini line and limb hair Skin rejuvenation shrink pores and whiten skin Operating procedures 1 Shaving 1 Before the op...

Page 23: ...utes and lip hair is about 3 minutes for reference only specific time should be based on hair changes and size of hair removal site Confirmed The clinical end point was observed when the hair follicle...

Page 24: ...pidermis 7 It is normal to feel hot pain in the pigmented area of the epidermis during treatment 8 Protect hair in non treated areas 9 The continuous working time of the instrument is too long it shou...

Page 25: ...Patients with hypertension and diabetes 4 Customers with light allergies 5 People who have been working under strong light for a long time and have recently sunburned skin 6 Customers with short term...

Page 26: ...ghtly warm a few parts will have redness and edema Late reaction the hair protrudes from the hair follicle the longer the hair color the lighter the longer the speed the slower the longer the thinner...