Sensor Replacement
4.
Sensor Replacement
(for oxygen sensors refer to section 4.1)
-
Screw off sensor cap and withdraw the old sensor
-
Remove the shorting socket from the new sensor
-
Plug the new sensor into the socket on the FMK 9002
Be sure to align the orientation guide properly before pushing the sensor into place.
-
Write down the concentration value CE stated on the sensor
-
Screw in the sensor cap
-
Perform calibration procedure
4.1
Replacing the Oxygen Sensor
-
Screw off sensor cap and withdraw the old sensor
-
Plug the new sensor into the socket on the FMK 9002
-
Be sure to align the orientation guides properly before pushing the sensor into place.
-
Screw in the sensor cap
-
Perform calibration procedure