MST FMK 9002 Manual To Operations Download Page 3

1.

Parts Identification

Functional Elements

1

Gas Sensor

  2

Sensor Cap

  3

4 ... 20 mA  Interface Connection

  4

Wall Mounting Support

  5

Digital Gas Concentration Display

Operational Controls for Sensor Calibration

  6

Zero Adjustment Potentiometer

  7

Span Adjustment Potentiometer

Operational Elements for Instrument Functions and  4 . . . 20 mA Interface Test

  

8

Selector Switch for Operating Modes

  9

LED Indication for Operating Modes

10

Push Button for  4 mA Simulation and Zero Adjustment of Sensor 

11

Push Button for 20 mA Simulation and Sensor Sensitivity

12

Test Plug for Alarm Simulation

3

2

1

9

8

11

10

4

6

7

12

5

FMK 

9002 9002.0000.61.0E 05/04

page 3 

Parts Identification

Summary of Contents for FMK 9002

Page 1: ...ithout notice Print No 9002 0000 61 0D 05 2004 Guide to Operation Remote Sensor Head FMK 9002 MST Technology GmbH Benediktstr 1 82069 Hohensch ftlarn Germany Fon 49 0 8178 8690 Fax 49 0 8178 8220 www...

Page 2: ...Alarms at the Central Controller FMS 8700 5 3 3 Operating Modes 5 3 4 Functional Test and Scaling External Control Systems 6 3 4 1 4 mA Simulation Zero 6 3 4 2 20 mA Simulation full scale 6 4 Sensor...

Page 3: ...meter 7 Span Adjustment Potentiometer Operational Elements for Instrument Functions and 4 20 mA Interface Test 8 Selector Switch for Operating Modes 9 LED Indication for Operating Modes 10 Push Button...

Page 4: ...K 9002 Connect the two wires of the cable fixed to the FMK 9002 with the two wire cable leading to the Central Controller FMS 8700 or FMS 8710 If the FMK 9002 is connected to other 4 20 mA Central Con...

Page 5: ...700 To prevent that alarm relays are activated and external alarms are released during electrical or gas technical checks the EXT ALARMS key switch at the Central Controller FMS 8700 must be set to th...

Page 6: ...he 4 20 mA loop is 4 mA Zero is indicated at the FMK 9002 and FMS 8700 8710 s displays Alarm is not activated at the FMS 8700 or FMS 8710 except with instruments for oxygen deficiency 3 4 2 20 mA Simu...

Page 7: ...sure to align the orientation guide properly before pushing the sensor into place Write down the concentration value CE stated on the sensor Screw in the sensor cap Perform calibration procedure 4 1...

Page 8: ...r oxygen Place the calibration cap P N 8600 4002 over the sensor and start the calibration gas flow flow volume approx 18 l h or 300 ccm min Allow the reading on the digital display to stabilize Turn...

Page 9: ...e performed in gas free atmosphere Selected FMK 9002 operating mode is calibrating Zero adjustment according to section 5 1 Press and hold button 20 mA CE Value calibration current will be fed Turn th...

Page 10: ...__ Sensor Adaptor standard 9002 0150 Sensor Cap standard 9002 0151 Test Plug package of ten 5 STEST 01K Calibration Cap standard 8600 4200 Wall Mounting Support 8601 0040 Guide to Operation German 900...

Page 11: ...0 Cl2 Chlorine 0 5 00 ppm 9012 5301 CO Carbon Monoxide 0 1000 ppm 9002 5400 COCl2 Phosgene 0 1 00 ppm 9012 6600 F2 Fluorine 0 5 00 ppm 9012 6400 GeH4 Germane 0 5 0 ppm 9012 6900 H2 Hydrogen 0 0 500 90...

Page 12: ...ib I BVS Nr 92 C 1003 Approved to meet the Intrinsic Safety Standard EEx ib IIC T6 BVS Nr 92 C 2002 Approved to meet the European Safety Standards EN 50014 and EN 50020 Copies of these approvals are...

Page 13: ...FMK 9002 9002 0000 61 0E 05 04 page 13 Certificates...

Page 14: ...FMK 9002 9002 0000 61 0E 05 04 page 14 Certificates...

Page 15: ...FMK 9002 9002 0000 61 0E 05 04 page 15 Certificates...

Page 16: ...FMK 9002 9002 0000 61 0E 05 04 page 16 Certificates...

Page 17: ...FMK 9002 9002 0000 61 0E 05 04 page 17 Certificates...

Page 18: ...FMK 9002 9002 0000 61 0E 05 04 page 18 Certificates...

Reviews: