9.3
Sensor Element
SPECIFICATIONS - Sensor Cartridge (SC2/MC2) / Sensor element
O
rd
e
ri
n
g N
o
.
M
e
asu
ri
n
g
ran
ge
A
cc
u
rac
y
R
e
so
lu
ti
o
n
R
e
p
e
atab
ili
ty
t90
Ti
me
Ze
ro
-p
o
in
t
var
iati
o
n
D
ri
ft
in
ai
r
Temper
atu
re
ran
ge
H
u
mi
d
ity
ran
ge
(
n
o
n
-
co
n
d
e
n
si
n
g)
Li
fe
ti
me
1
in
ai
r
R
e
lati
ve
g
as
d
e
n
si
ty
M
o
u
n
ti
n
g
h
e
ig
h
t
C
al
ib
rati
o
n
in
te
rv
al
1
Ze
ro
G
ai
n
MC2-/
SC2-
ppm
± %
sig.
ppm <± %
sig.
≤ sec. ±ppm < % signal/
month
°C
% r. F.
>
months
Air = 1 (m)
Month
E1125-AX
0-100
5
2
10
40
10
1
1
-30 / +50 15-90
24
0.59
Ceiling
12
E1125-BX
0-300
2
4
10
40
50
1
2
-30 / +50 15-90
24
0.59
Ceiling
12
E1125-CX
0-500
3
4
10
40
50
1
2
-30 / +50 15-90
24
0.59
Ceiling
12
E1125-DX
0-1000 3
4
10
40
50
1
2
-30 / +50 15-90
24
0.59
Ceiling
12
E1125-EX
0-5000 2
5
10
40
100
1
2
-30 / +50 15-90
24
0.59
Ceiling
12
E1193-BX
0-5
n.D. 0.1
2
40
0.2
1
2
-20 / +50 15-90
24
2.4
Floor
6
E1193-CX
0-10
n.D. 0.1
2
40
0.2
1
2
-20 / +50 15-90
24
2.4
Floor
6
E1193-DX
0-20
n.D. 0.1
2
40
0.2
1
2
-20 / +50 15-90
24
2.4
Floor
6
E1189-CX
0-200
n.D. 0.3
1
60
4
1
3
-20/ +50
15-90
24
0.97
1.5-1.8 12
E1185-BX
0-10
0,5 0.01 2
60
0.2
1
2
-10/ +45
15-90
36
1.09
Floor
6
E1110-BX
0-100
3
0.5
5
10
4
0.4
0.4
-15 /+50
10-95
72
0.97
1.5-1.8 12
E1110-CX
0-150
2
0.5
5
10
4
0.4
0.4
-15 /+50
10-95
72
0.97
1.5-1.8 12
E1110-EX
0-250
2
0.5
5
10
4
0.4
0.4
-15 /+50
10-95
72
0.97
1.5-1.8 12
E1110-FX
0-300
2
0.5
5
10
4
0.4
0.4
-15 /+50
10-95
72
0.97
1.5-1.8 12
E1110-HX
0-500
2
0.5
5
10
4
0.4
0.4
-15 /+50
10-95
72
0.97
1.5-1.8 12
E1190-AX
0-5
n.D. 0.05 5
30
0.15
1
2
-10/+45
15-90
24
1.66
Floor
12
E1190-BX
0-10
n.D. 0.05 5
30
0.15
1
2
-10/+45
15-90
24
1.66
Floor
12
E1187-AX
0-5
n.D. 0.03 2
25
0.06
1
2
-10/+45
15-90
24
1.18
Floor
12
E1196-BX
0-20
2
0.2
2
30
0.1
1
2
-10/+45
15-90
24
2.26
Floor
12
E1197-AX
0-50
3
0.1
2
30
0.1
1
2
-10/+50
15-90
24
1.19
Floor
12
E1197-BX
0-100
2
0.1
2
40
0.1
1
2
-10/+50
15-90
24
1.19
Floor
12
E1197-CX
0-200
2
0.1
2
40
0.1
1
2
-10/+50
15-90
24
1.19
Floor
12
E1197-DX
0-500
n.D. 0.1
2
40
0.1
1
2
-10/+50
15-90
24
1.19
Floor
12
E1188-AX
0-50
n.D. 0.1
2
60
1
2
2
-10/+45
15-90
24
1.16
Floor
12
E1130-AX
0-10
5
0.1
2
25
0.2
1
2
-20/ +50
15-90
24
1.59
1.5-1.8 12
E1130-BX
0-20
5
0.1
2
25
0.2
1
2
-20 / +50 15-90
24
1.59
1.5-1.8 12
E1130-CX
0-30
5
0.1
2
25
0.2
1
2
-20/ +50
15-90
24
1.59
1.5-1.8 12
E1130-DX
0-500
n.D. 2
2
40
0,2
1
2
-20/ +50
15-90
24
1.59
1,5-1.8 12
E1129-CX
0-100
2
0.5
2
10
3
1
2
-10 / +50 15-90
12
1.04
1.5-1.8 12
E1129-DX
0-1000 20
2
2
30
3
1
2
-10/ +50
15-90
12
1.04
1.5-1.8 12
Vol %
E1195-A
2/3/5/7
0-25
2
0.05 --
15
--
--
0.3
-10/50
5-95
24 / 36
60 / 84
1.5-1.8 24