8
6056 Rev B
7.
Pull the elastic strap across the fluid bag and secure by affixing the I.V. Bag Strap.
8.
Turn the unit “on” by pressing the Power Switch. The switch and the LED indicators on
the bottom of the Heater Plate will illuminate when energized. (Ref: Figure 1)
.
If fluid spills inside the Heater Plate or Power Box, unplug the unit to prevent an
electrical shock hazard and refer to the Cleaning Instructions (Section 5.0).
5.0
CLEANING INSTRUCTIONS
1.
Turn the main power “off” and unplug the
T° HOLD
®
from the power source.
2.
Wipe down the Heater Plate with a damp cloth using a general hospital cleaner
(Isopropyl Alcohol) as needed, following the cleaner manufacturer’s instructions for use.
3.
If excessive fluid enters the Heater Plate, disconnect power to the unit and refer to
qualified Bio- Med personnel for service.
4.
Contact Medical Solutions, Inc., at 800-435-7600 for further instruction.
6.0
TROUBLESHOOTING
1.
If the
T° HOLD
®
fails to warm, check to ensure that:
a)
The power cord is plugged into an approved hospital grade outlet.
b)
The power switch is turned on at the Power Box with the Power Indicator Light
illuminated. (Ref: Figure 6A).
c)
The green LED Indicators on bottom of the Heater Plate are illuminated.
(Ref: Figure 6B).
2.
If the Power Indicator Light is not illuminated please have a Bio Medical technician
verify that the two fuses on the Power Box are functional. (Ref: Figure 3A).
3.
If the power is “on” and the
T° HOLD
®
fails to warm, please contact Medical
Solutions, Inc. at 800-435-7600.
Figure 6A
Figure 6B
FIGURE 6- POWER INDICATOR LIGHT AND LED INDICATORS
Power Switch
LED Lights