Page 34
MSI-4300 Porta-Weigh Plus/7 Key
•
User Guide
M E A S U R E M E N T S Y S T E M S I N T E R N A T I O N A L
When the desired mode "ENABLED", or "DISABLED" is
displayed push ENTER.
Caution
: Disabling Auto Zero Maintenance will degrade temperature
and drift performance of the Porta-Weigh Plus.
Disabling the AZM is only intended for certification testing.
Under no other circumstances should the scale be used with the AZM
disabled.
When calibration is finished:
Seal the Calibration Port.
Calibration cycle:
Measurement Systems International recommends that the Porta-Weigh
Plus is calibrated at least once a year by a qualified scale technician
with certified calibration standards.
EIA-232-D SERIAL OUTPUT OPTION
Function:
The EIA-232 output is used in conjunction with the Infrared Remote
Control to output weight and total data to a printer or can be used for
two-way communications with a computer.
Electrical Conformance:
The electrical characteristics of the serial output is configured to
conform to the EIA Standard EIA-232-D. Cable connections include
RXD, TXD, Ground, CTS, and RTS. (CTS / RTS handshaking is not
currently supported, consult factory)
Configuration:
The Porta-Weigh Plus serial port option is factory configured only.
Standard data configuration is:
Baud Rate:
9600
Data Bits:
8
Parity:
None
Start Bits:
1
Stop Bits:
1
By customer order, the standard configuration can easily be changed.
Available configurations include any combination of the following
settings: