17
System Concept
GB
SUPREMATouch
INT Terminals (Internal Power Supply, 24 VDC 250 W)
• Connection for voltage supply from an internal rack power supply or an external power supply
unit.
• Power supplied to all rack units and the sensors.
• Internal power supply unit (MSP module) has a supply voltage input of 85–265 VAC
(47–63 Hz) or 120–330 VDC.
• If the rack power supply unit cannot supply enough power, the sensors, modules or relays must
be supplied by external power supply units.
• The internal rack power supply can be omitted if, because of a high power requirement or a
redundant design, the power is being supplied by an external power supply via the INT termi-
nals.
• Maximum supply current of 20 A.
BAT Terminals (Backup Battery Power Supply)
• Backup battery power supply for all units of a rack (21 –28 VDC).
• In the case of failure of internal and/or external power supply, the backup battery power supply
will supply power to the system.
• Maximum supply current of 20 A.
Power Supply Plans
All of the system cards and the sensors can be supplied from each of the 3 pairs of terminals
connections. A voltage changeover switch is provided on each system card, which ensures that
only one of the voltages being applied is accepted. Various power supply plans are available to
suit the number and type of sensors and/or the required degree of redundancy in the power
supply.
If the internal rack power supply unit is not sufficient to power all the sensors, an external unit must
be provided. The internal unit must then be disconnected. A redundant power supply is then
provided by external units via the BAT or INT terminals.
Supply Plan A: Internal Power Supply Unit
All of the units of the system and the sensors are supplied by the rack power supply unit (INT
terminals). This variant is used when power supply redundancy is not required and the power
which can be supplied by the unit installed in the rack (250 W) is sufficient to supply all of the rack
modules and the connected sensors.
Supply Plan B: External Power Supply Unit
All system modules housed in the rack and the sensors are supplied by the external power supply
unit (EXT terminals). This variant is used when power supply redundancy is not required and the
power of the unit installed in the rack (250 W) is not sufficient to supply all of the system modules
and the connected sensors. A maximum of 20 A can be supplied across the terminals (480 W
system power).
Supply Plan C: Internal Power Supply Unit + Battery
All units of the system and the sensors are supplied by the rack power supply unit (INT terminals)
or by the backup power supply (BAT terminals). This variant is used when there must be redun-
dancy in the power supply and the power of the unit installed in the rack (250 W) is sufficient for
all rack modules and the connected sensors.
Supply Plan D: External Power Supply Unit + Battery
All modules of the system and the sensors are supplied by the external power supply unit
(EXT terminals) or by the backup power supply (BAT terminals). This variant is used when the
power supply must be redundant and the power that can be supplied by the unit installed in the
rack is not sufficient to supply the system modules and the connected sensors. A maximum of
20 A can be supplied across the terminals (480 W system power).
2.8
Safety Concept
The individual functional modules are connected to each other by a CAN bus. The CAN bus is
designed to be virtually error-proof. Every module can detect errors on the bus and handle them
appropriately. The probability of an undiscovered communications error on the bus is 4.7 * 10
-14
.
Error statuses on the CAN bus are indicated on the D OPERATION unit (MDO module).
Summary of Contents for SUPREMATouch MCP 20
Page 1: ...MSAsafety com Operating Manual SUPREMATouch Fire and Gas Warning Unit Order No 10121863 05...
Page 8: ...MSAsafety com User Instruction Manual SUPREMATouch Fire and Gas Warning Unit...
Page 72: ...MSAsafety com Service and Maintenance Guide SUPREMATouch Fire and Gas Warning Unit...
Page 232: ...232 Dimensions GB SUPREMATouch 18 Dimensions 18 1 Rack...
Page 233: ...233 Dimensions GB SUPREMATouch 18 2 Rail mounted Modules MRO8 TS Module MRO16 TS Module...
Page 235: ...235 Dimensions GB SUPREMATouch MRO20 8 TS Module 1 3 2 69 90...
Page 236: ...236 Dimensions GB SUPREMATouch MRO20 16 TS Module 2 5 2 64 73 relay dependent 90...
Page 237: ...237 Dimensions GB SUPREMATouch MRC TS Module MGT40 TS Module...
Page 238: ...238 Dimensions GB SUPREMATouch MHD TS Module MAT TS Module...
Page 241: ...For local MSA contacts please visit us at MSAsafety com Because every life has a purpose...