SingleLine SCOUT
6
Installation
GB
3
Installation
Disassembling SingleLine Gauge
For disassembling the SingleLine gauge, a Torx spanner and a small screw driver are required.
CAUTION!
The screws are not reusable and MUST be replaced with new ones after each disassembly.
(1)
Unscrew the back cover of the SingleLine gauge with a
Torx T10 spanner.
(2)
Lift the back cover.
(3)
Remove the U-clip with a small screw driver.
(4)
Pull the SingleLine manifold out.