6.8.
Automatic zero-point setting
Touch the display the press the green ON button. The analyzer will per-
form a self-check and start zeroing.
During zeroing the probe must be in ambient air (not in the stack)!
During zeroing there is a blue zeroing bar (top of display) visible.
Once zeroing is completed, the blue bar will disappear and the
DELTA
smart
is ready for measurement.
The analyzer will display a defect sensor if any of the installed sensors is
not ready after zeroing.
Repeating the zeroing
Zeroing can be repeated at any time as long as the probe is not inside the
stack and at ambient air.
In the main menu „Measurement“
you select the „ze-
roing“ button.