MRC UT-FLU5B Operation Manual Download Page 14

8

dsDNA detection.

2.1 dsDNA detection

Click the dsDNA icon to enter the following interface (Figure 5)

Figure 5 Sample detection

In Figure 5:

Standard curve: Select the established standard curve, and the curve information will be

displayed in the lower left corner.

Sample detection: After entering, the fluorescence value of the fluorescent sample can be

detected, and the sample concentration can be calculated according to the selected standard
curve.

Note:

a. When the 0.5mL sample tube is used for detection by the fluorometer, the volume of

the sample to be detected is 200uL. When the sample volume added during the detection is
less than 200uL, it needs to be supplemented to 200uL by the diluent. The concentration of
the sample detected at this time is the concentration of the sample in the tube, while The
instrument calculates the original sample concentration (that is, the actual concentration of the
added test sample) from the added sample volume and the diluent volume.

Actual sample volume
added

Standard

curve

selection

Summary of Contents for UT-FLU5B

Page 1: ...UT FLU5B FLUOROMETER OPERATION MANUAL 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC 12 22...

Page 2: ...f 1 D ONA Oil co Protein lo _ Cuetoni Fluometer Sett Ina...

Page 3: ...carefully before operating the instrument Please keep the manual in a safe place so that you can read it quickly if you encounter a problem Unpacking Inspection When the user unpacks the instrument f...

Page 4: ...t all precautions contained therein This instrument is an ordinary device that complies with the GB4793 1 standard and this instrument is the product for indoor use Please read this manual carefully b...

Page 5: ...e power cord Make sure that the plug is fully inserted into the socket when plugging in and do not pull the power cord hard when unplugging Turn off the power when it stops working When the instrument...

Page 6: ...repair or replace instruments that prove to be defective Warranted products must be sent by the user to the maintenance department determined by the company The shipping cost of the instrument from t...

Page 7: ...antibodies nucleic acids hormones steroids thyroid hormones phthalohormones drugs and microorganisms This fluorometer is based on fluorescence immunoassay technology which detects the luminescence int...

Page 8: ...n wavelength Emission wavelength UT FLU5B I Fluorimeter Blue LED 460 20nm 525nm Red LED 625 20nm 690nm UT FLU5B II Fluorimeter Blue LED 460 20nm 525nm UV LED 365 20nm 470nm UT FLU5B III Fluorimeter Bl...

Page 9: ...the preparations before it is turned on When using the instrument for the first time you should familiarize yourself with the contents of this chapter before turning it on Structure diagram Figure 1...

Page 10: ...ackage and check after unpacking in the presence of agents and staff According to the packing list check whether all components are complete Carefully check the appearance of all devices to see if the...

Page 11: ...fluorescent sample into the PCR tube holder 3 Cover the sample cover and click detecting in the interface to perform sample detection Figure 2 Sample placement In order to ensure the detection effect...

Page 12: ...the instrument starts and enters the self test interface Figure 3 Boot interface After the self check is completed enter the main interface of the instrument The following sections describe each inte...

Page 13: ...concentration analysis The other 5 icons after entering can inject standard curve establishment calibration curve and sample detection The specific functions of these five icons are exactly the same a...

Page 14: ...rding to the selected standard curve Note a When the 0 5mL sample tube is used for detection by the fluorometer the volume of the sample to be detected is 200uL When the sample volume added during the...

Page 15: ...trument the display here is blank Click the blank area to pop up a drop down menu in which NEW is the button for creating a new curve When there are multiple curves you can also click the curve name a...

Page 16: ...unnecessary or mishandled fluorescence values Methods for establishing a standard curve a Prepare a series of standard samples 5 9 standard samples b Create a new standard curve and enter the standar...

Page 17: ...ibrate the curve under the condition of instrument drift which can save the sample and operation time for establishing the standard curve while improving the sample detection accuracy The calibration...

Page 18: ...ire curve Three point calibration Two high medium and low concentration sample points that can basically cover the concentration range of the entire curve are used and it is better that the three samp...

Page 19: ...ng icon to enter the next level interface Note The instrument is configured with 2 channels so the channel needs to be selected during detection The instrument configuration channels vary according to...

Page 20: ...interface can detect samples print data etc Note Printing requires an external printer to be connected Figure 12 Data record The data recording interface saves all test data which can be viewed delet...

Page 21: ...nu to enter the following system settings interface Figure 4 13 System setting 4 1 Language Click the Language icon a language selection dialog box will pop up select the desired language and click OK...

Page 22: ...ck Set Time to enter the time setting interface shown in Figure 4 17 Click the 10 27 53 time area to modify the time and click 2020 07 04 to modify the date 4 3 Brightness adjustment Click the Screen...

Page 23: ...tware in the root directory of the U disk then insert the U disk into the machine and then click the software update icon to enter the following interface Figure 4 17 Software update Click upper softw...

Page 24: ...of the machine Figure 4 18 Instrument information 4 6 Maintenance The maintenance function is only used in the production and maintenance links The interface is not involved when the user uses it and...

Page 25: ...an be turned on or off in the interface In the Auto Print mode the current test result will be automatically printed after the test is completed Figure 4 19 Print setting Note The thermal printer is a...

Page 26: ...n it is found that the instrument is abnormal or can not work normally you should contact the manufacturer in time and users should not disassemble it at will 3 The power switch should not be turned o...

Page 27: ...roperly connected The module contact is abnormal 1 Check whether the power connection is normal 2 Check whether the module is loose if so contact the supplier or manufacturer 3 The measurement is unst...

Page 28: ...NO Name Specification Unit Quantity Remarks 1 Power adapter 15V Piece 1 With ground 2 U disk 8G Piece 1 Data import export 3 0 5mLPCR tube socket or 0 2mLPCR tube socket Piece 1 Put it in the instrum...

Reviews: