19
Operation stops with
alarm:
Display shows
Circuit is interrupted
- Verify that the running buffer is correct.
- Verify the all cables are attached correctly
- Turn the Power switch off and on again;
restart application.
- If you cannot restart the instrument, turn off
the power, disconnect the power cord from
the outlet, and contact Technical Service.
Operation stops with
alarm:
Display shows
Circuit is interrupted
- Verify that the running buffer is correct.
- Verify the all cables are attached correctly
- Turn the Power switch off and on again;
restart application.
- If you cannot restart the instrument, turn off
the power, disconnect the power cord from
the outlet, and contact Technical Service.
Operation stops with
alarm:
Display shows
Ground leak detected
during run
Check the electrophoresis system for
improper grounding. Restart the power
supply by turning the Power
switch off and on.
Operation stops with
alarm:
Display shows
Power supply is
overheating
- Turn off power supply. Check for sufficient
airflow around the power supply fan. After
cooling down, restart the power supply by
turning the Power switch to the on
position.
- If you cannot restart the instrument, turn off
the power, disconnect the power cord
from the outlet, and contact Technical
Service.
Operation stops with
alarm:
Display shows
and countdown
Power supply is getting
restart ON
- Verify that the electrophoresis condition
and system is correct.
- Pay attention to the own safety.