If you are using a 6.8mm lens, a person should stand 3-5m from the camera.
•
If you are using a 19mm lens, a person should stand 6-10m from the camera.
•
The temperature indicator will give the surface temperature of the individual
with the hottest temperature in the field of view, or the configured region of
interest.
•
When you receive an alert for an individual, check more thoroughly with a
thermometer.
•
The IR-CAM-FEVER is affected by the ambient temperature of its surroundings and it
provides relative temperature readings, as opposed to a thermometer which gives
absolute temperature readings.
•
While the baseline may change, the hightened temperature reading of a feverish
individual will give an alert.
•
Note:
The skin temperature of a person may be affected by; extreme
temperatures such as extreme cold or standing near a heat source, drinking
significant quantities of alcolhol, taking temperature altering drugs, or strenuous
exercise.