The air flow rate is too high or too low
(FID).
10)
The hydrogen or air is contaminated.
11)
The water condenses in FID.
12)
The detector cable is in poor contact.
13)
The detector’s insulation turns smaller
(the ionization detector).
14)
The electrode, the nozzle or the base of
the detector is contaminated.
8)
Re-adjust the hydrogen flow rate.
9)
Re-regulate the air flow rate.
10)
Replace both the hydrogen filter and
the air filter.
11)
Remove the water by increasing the FID
temperature. (14) Replace the cable or
repair it.
12)
Clean the detector insulator.
13)
Clean the detector.
18.
Periodical baseline
1)
The detector’s temperature control is
deficient.
2)
The control of the chromatography
column oven is deficient.
3)
The carrier gas flow is not set properly.
4)
The pressure of the gas flow is too low.
5)
Air and hydrogen are not adjusted well
(FID).
1)
Check the platinum insulator and
improve the control precision.
2)
Check the platinum insulator and
improve the control precision
3)
Adjust the flow rate of the carrier gas
4)
Replace the carrier gas cylinder.
5)
Regulate the hydrogen and air flow.
19.
One-way baseline 1)
There is a significant increase or
1)
Stabilize the detector temperature. If