- 3 -
(6).The elevation connection using the transmission technique of the linear
bearing brings the flexible elevating operation.
(7).All information can be clearly displayed on the screen. The preset
temperature, actual temperature and heating time can be displayed
simultaneously on the screen.
(8).The safe operation is ensured because the electric circuit is completely
isolated with the extraction cell.
(9).With the unique air space insulation technique, the temperature of
instrument’s case can be kept normal.
(10). Compared with the traditional soxhlet extraction, its speed and
efficiency are better.
3. Working Condition
(1). Input voltage: 220V 50Hz
;
air break switch and leakage circuit breaker
are required.
(2). The apparatus should be placed where water, drain basin, electric pocket
are available.
(3). Power supply should meet the requirements of the apparatus to avoid
electric over stress; independent switch, safety device and reliable grounding
are required.
(4). The apparatus should be placed away from big electric equipment to
avoid strong magnetic field.
(5). Ensure the good ventilation of the laboratory.