2-2 Electrical Specifications
( 23 5 )
Watt (AC, true power),
current mode from direct input
Range
Resolution
Accuracy
6,000 Watt
1 Watt
( 1.5% + 5 d )
* Accuracy are specified under the following
conditions :
a) AC input current is 0.4 ACA & 10 ACA.
b) AC input voltage is within 110 V 15 % and
220V 15%.
c) ACA, ACV input signal is sine wave, 50/60 Hz.
d) Power factor 0.8.
* ACA, ACV frequency response is from 40 to 400 Hz.
* Max. volt & current input signal value :
Volt input : Max. AC 600V, Current input : Max. AC 10 A
Watt (AC, true power),
current input cooperate with inductive probe or CT
Range
Resolution
9,999 Watt
1 Watt
99.99 KW
0.01 KW
999.9 KW
0.1 kW
* Accuracy will be same as the abve " Direct Current
Input Mode " but plus the accuray value of Current
Transformer ( CT ) or the acccuracy value of Inductive
Current Probe.
*
Input current should obey :
Inductive Probe - 20 ACA.
CT 100/5 A - 8 ACA.
CT 1000/5 A - 80 ACA.
3