Setting the Temperature (Continuous mode) 

Put the front switch on CONTINUES 

Turn on the temperature controller/power switch. The temperature displays light up 
and after a few seconds, the upper display shows the present value (PV) temperature. 
The lower display shows the set value (SV) or set temperature. 

   

Press 

 to increase temperature gradually. 

        Press 

 to decrease temperature gradually. 

 
 
Setting the Temperature (timer mode) 
 

Put the front switch on TIMER 

Turn on the temperature controller/power switch. The temperature displays light up 
and after a few seconds, the upper display shows the present value (PV) temperature. 
The lower display shows the set value (SV) or set temperature. 

   

Press 

 to increase temperature gradually. 

        Press 

 to decrease temperature gradually. 

 

Press number 

 twice. PV display 

, SV display 1.0 

Press

 

 to increase or decrease time gradually. Setting time upon what you 

need.  
The timer will start from the temperature reach setting. 
 
For example:  
1.5 = 90 seconds (1.5 minutes), 1.0 = 60 seconds (1.0 minutes) 0.5 = 30 seconds. 

Press number 

 once. PV display 

 , and SV display 1.0 

 It mean the time 

will counting down 1 min after temperature reach to the setpoint.  

 
 
User Calibration (only if needed) 

Each unit is calibrated in the factory before shipment. The user can still modify the 
calibration in the field.  
 
There are two parameters: Offset Low (OFTL) and Offset High (OFTH) for 
adjustment to correct an error in the process value.  
 

Press and hold the 

 key until the setup 

 Menu page is obtained.  

Then, press 

 key to scroll to the parameter

.   

Press 

 key several times to scroll to the parameter OFTL to calibration low 

value. Send your low signal to the sensor input of the controller, then press and 

release the

 key.  If the process value (the upper display) is different from the input 

signal, the user can use 

 and 

 to change the OFTL value (the lower display) 

until the process value is equal to the value the user needs. Press and hold the 

key 

Reviews: