20
7.2.2 The use of commercially available plastic syringe
7.2.2.1 Sampling
1.
Lower the sample discharging lever to the bottom.
2.
Push the intake lever stopper upward to lock it.
3.
Remove the screw stopper with a coin. Do not lose
the screw stopper.
4.
After fixing optional grommet (98-561-0005), squeeze
in the syringe adapter (98-581-0057).
5.
Leave dipped the tip of sampling nozzle in the waste
bottle.
6.
Inject sample liquid into the cell with the syringe.
7.
When the cell is filled with sample, leave the syringe as
it is.
Caution!
When filling the sample with a syringe, push it as slowly as possible.
If it is abruptly pushed, the unit can be broken as the result of
detachment of the sample discharging lever.
7.2.2.2 Start measurement
Only when
is displayed (Stability:Auto), start measurement with [ok/meas.] key.
7.2.2.3 Hold measurement data
When
is displayed (Stability:Auto), the value will automatically turn to get reversed when the reading
stabilizes.
If
is not displayed (Stability:Manu), wait until the reading stabilizes, and press [ok/meas.] key. The
reading will be held in reversed mode.
7.2.2.4 Saving measurement data
When is displayed (memory in: Auto), the reading will be automatically saved when retained.
If is not displayed (memory in: Manu), press [ok/meas.] key while the reading is held. The data at
this point will be saved.