use of larger volume of the angle rotor or swing-out rotor, you can choose a little bit of the

slows, the other can choose a little faster file Bit. Generally recommended to use 5 files;

Finally press "Enter" to confirm.
7) When the "Short" button is pressed, the machine centrifuges, and when the mobile phone

is loosened, centrifugation stops.

Remind: each step is completed, should immediately press the "Enter" to be confirmed.

8) Finally, check the parameters on the screen confirm correct, press the "Start" " to run.
9) The running time countdown to "0", the centrifuge will automatically stop, the buzzer

ring, when the speed is 0, press the "Stop / Open" button, you can open the lid. During the

event of an emergency, press the "Stop / Open" button, the buzzer does not ring when the

centrifuge stops running.

Note: Do not open frequently, the proposed interval of 10 seconds to avoid

damage to components

10) Open, take out the tube carefully, complete the separation process.
11) turn off the power switch, cut off the centrifuge power.

VIII

Troubles & Solutions

Troubles

Cause

Solution

No display when power

plugged in

No power

Power check

Fuse blown

Check & replace

Cable connection loose

Check & reconnect

Large vibration

Tube No Symmetrically placed Check & adjust

Tube broken

Check & replace

rotor not tightened

Check & adjust

Partial damage to shock

Check & replace

Display 0000

Start no

running

Circuit board or transformer

damaged

replace

control system connector is

loose

Re-tighten

Summary of Contents for CENVCC-3

Page 1: ...ENVCC 3 Table Vacuum Centrifugal concentrator Instruction Using Manual Please read this manual carefully before use 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC 8 22 ...

Page 2: ...VACUUM CONCENTRATOR SPECIFICATION 2 Ⅳ ROTOR SPECIFICATION 4 Ⅴ PRECAUTION 5 Ⅵ APPEARANCE AND STRUCTURE 6 Ⅶ WORKING PROCEDURE 7 VIII TROUBLES SOLUTIONS 8 IX INSTALLATION 9 X MAINTENANCE 10 XI WARRANTY 11 XII CERTIFICATE OF CONFORMITY 12 XIII PACKING LIST 13 ...

Page 3: ...entrator is a method in which the sample is continuously pumped by a vacuum pump while being centrifuged or even close to vacuum so that the solvent in the sample is quickly evaporated the sample system is rapidly concentrated or dried and the cold trap and the filtration method can be effectively used which collected evaporated solvent This model is suitable for RNA DNA in the fields of molecular...

Page 4: ... the speed range allows It is strictly prohibit to work over speed 5 No unbalance working centrifuge tube must be placed symmetrically no single tube working the tube solution must be uniform to ensure balanced operation 6 No high speed working without rotor 7 Remove the rotor must use the special handle which together with the machine or use non metallic tools gently tap the rotor first repeated ...

Page 5: ...nt or rotor disabled for three month please running at low speed 10min then allow to work on high speed 12 To prevent damage to the instrument and personal safety No mixed using other company Rotors 13 To prevent the machine and spare parts corrode Do Not place a container with liquid on the instrument 14 Please clean the machine and close the power switch and power plug after using Ⅵ Appearance a...

Page 6: ... Acceleration gear 7 Deceleration gear 8 Program group number 9 Setting key 10 increasing key 11 instantaneous centrifugal key 12 Running indicator 13 Start key 14 Confirmation key 15 Reduction key 16 Speed Centrifugal Force Mode Switching Key 17 Stop indicator 18 Stop key open key ...

Page 7: ...ow will display E3 6 set the rotor number speed time and other operating parameters 1 Set the speed press the Set key move the cursor to the speed setting position when the speed digital flicker that is enter the speed setting then press the increase key or reduce key to set the speed of this operation after setting press the confirmation key to confirm the parameters Special reminder The horizont...

Page 8: ...Stop Open button you can open the lid During the event of an emergency press the Stop Open button the buzzer does not ring when the centrifuge stops running Note Do not open frequently the proposed interval of 10 seconds to avoid damage to components 10 Open take out the tube carefully complete the separation process 11 turn off the power switch cut off the centrifuge power VIII Troubles Solutions...

Page 9: ...tor stall E8 Communication error E9 System error EA Low voltage EB Overcurrent protection EC Motor fault Hall state error ED Brake resistance Note The above alarm is displayed in the display window when appears the buzzer alarm for a long time short and alarm shutdown In the event of an alarm the machine will be correct display only when the fault is released If the fault is not released press the...

Page 10: ...to ensure that the four foot contact with the table B centrifuge around the gap with the wall should be greater than 16cm to ensure good ventilation C power exchange 220V 50Hz 10A or 110v 60Hz the wire cross sectional area of not less than 2 square millimeters the socket capacity of not less than 10A the protective grounding wire must be connected firmly X MAINTENANCE 1 After using such as the day...

Page 11: ...ity the following More than the max speed of the rotor used The rotor has scratches and bumps The rotor is seriously corroded by acid and alkali Unauthorized professional maintenance 3 without the permission of the Company non authorized professionals shall not open the machine all maintenance and adjustment must be carried out by professionals in accordance with the maintenance manual 4 the Compa...

Page 12: ...XII CERTIFICATE OF CONFORMITY Table Vacuum Concentrator CERTIFICATE OF CONFORMITY Max RPM 1800r min Max RCF 409 g Serial number This product has been tested and approved Inspectors ...

Page 13: ...ncentrator 1Set 二 Accessories tools 1 Power plug 1Pcs 2 Angle rotor 62 1 5ml 1Set 3 Vacuum Pump 1set 4 Cold Trap 1set 5 6 7 8 9 三 Document 1 Instruction manual 1pc This manual 2 Certificate of conformity 1pc This manual 3 Packing list 1pc This manual Packing Inspector July 19 2022 ...

Page 14: ...pply 220V 22V power 500W Overall dimension 665 four hundred and ten 420 long wide Height in mm Pipe joint size φ 10mm Temperature setting range 40 normal temperature Temperature control accuracy 1 Volume of freezing chamber 4L 2 The control interface of the cold trap is as follows Control panel easy to operate and use ...

Page 15: ...e for vacuum pump and lock it with clamps 3 Plug the power cord into the power socket 4 Turn on the power switch 5 Press the start button of the compressor to cool the compressor At this time the operation indicator lights up If you need to stop cooling press the stop button directly the compressor will also stop when the temperature reaches the set temperature The temperature control settings are...

Page 16: ... a large amount of effusion in the freezing cavity use a tool to unscrew the water drain valve on the left side of the instrument to discharge the effusion confirm that the seal is complete when tightening 7 Please turn off the power after use 8 The cold trap host provides one year warranty In case of non human damage to the product the company will provide free maintenance service within 1 year f...

Reviews: