09/ 2013
7.1 Direct voltage ranges
The input impedance is 10 MΩ.
Range
Resolution
Accuracy
Overload protection
320 mV
100 µV
± (0,5 % of r 2 digits)
600 V DC/ AC
3,2 V
1 mV
± (0,5 % of r 2 digits)
600 V DC/ AC
32 V
10 mV
± (0,5 % of r 2 digits)
600 V DC/ AC
320 V
100 mV
± (0,5 % of r 2 digits)
600 V DC/ AC
600 V
1 V
± (0,5 % of r 2 digits)
600 V DC/ AC
7.2 Alternating voltage ranges
The input impedance is 10 M
parallel 100pF. The measurement value is
arrived at by average sensing and is displayed as the RMS value.
Range
Resolution
Accuracy in 50 Hz - 400 Hz
frequency range
Overload protection
3,2 V
1 mV
± (1,5 % of r 5 digits)
in 40 Hz - 300 Hz frequency range for 3 V-range
600 V DC/ AC
32 V
10 mV
± (1,5 % of r 5 digits)
600 V DC/ AC
320 V
100 mV
± (1,5 % of r 5 digits)
600 V DC/ AC
600 V
1 V
± (1,5 % of r 5 digits)
600 V DC/ AC
7.3 Direct current ranges
Range
Resolution
Accuracy
Burden voltage
Overload
protection
320 µA
0,1 µA
± (1,0 % of reading
+ 2 digits) or 3200 µA
< 3 mV/ µA
600 V DC/ AC
3200 µA
1 µA
± (1,0 % of reading
+ 2 digits) or 3200 µA
< 3 mV/ µA
600 V DC/ AC
7.4 Resistance ranges
Overload protection for resistance measurements: 600 V
rms
.
Range
Resolution
Accuracy
Max. open circuit voltage
320 Ω
0,1 Ω
± (1,0 % of r 4 digits)
1,3 V
3,2 kΩ
1 Ω
± (0,8 % of r 2 digits)
1,3 V
32 kΩ
10 Ω
± (0,8 % of r 2 digits)
1,3 V
320 kΩ
100 Ω
± (0,8 % of r 2 digits)
1,3 V
3,2 MΩ
1 kΩ
± (0,8 % of r 2 digits)
1,3 V
32 MΩ
10 kΩ
± (0,8 % of r 5 digits)
1,3 V
7.5 Diode Testing
The accuracy indicated below is valid in the range of 0.4 V to 0.9 V.
Overload protection for diode testing: 600 V
rms
/ 600 V DC.
Range
Resolution
Accuracy
Maximum
current
Max. open circuit
voltage
1 mV
± (1,5 % of r 5 digits)
1,5 mA
3,3 V
7.6 Continuity Testing
The built-in buzzer sounds if the resistance R falls below 20 Ω.
8. Measuring with
01
8.1 Measurement preparation
The
MN 1
must be used and stored only at the indicated storage and
working temperatures, avoid exposure to continuous sunlight.
- Check the rated voltage and current indications on the test leads. The
safety test leads provided with the
01
correspond specifically
to the rated voltage and current of the
01
.
- Inspect the test leads for damaged insulation. If the insulation is damaged,
the test leads should be discarded immediately.