3-2
Receive Performance Check
: Tone Sensitivity
3.2
Tone Sensitivity
1. Connect the TEM cell and HP8920 or equivalent equipment as shown in Figure 3-1. below.
2. Put the pager into a TEM Cell, set up the the function switch to test frequency.
3. Set up the amplitude of the signal generator according to a known signal.
4. In this state, the HP8920 (or equivalent equipment) should be set up as follows.
a. QC2 signal should be sent to the pager by clicking “Send”, and make sure that the Alert
tone is OK.
b. Repeat step 4a., lowering the amplitude each time until 4 out of 5 Alerts are received.
Figure 3-1. Tone Sense Check Setup
NOTE:
Tone Sensitivity results may vary based on pager position and orientation in TEM cell.