TENSION CONTROL OF FOCUSING MOVEMENT
The focus tension is easily adjusted by using the adjustment wrench.
This wrench fits the tension collar found on the focusing controls,
between the knob and upright support. A clockwise turn of this collar
moves it toward the upright support and increases tension, while a
counter clockwise turn moves the collar toward the knob and decreases
tension (see illustration).
ADDING A DIGITAL CAMERA TO TOP CAMERA PORT
The top port of your SM105-T microscope allows the attachment of
a digital camera, such as the Moticam. The purchase of an optional
C-mount camera lens is necessary for the conversion (MA15604). DO
NOT PLUG INTO USB PORT OF MICROSCOPE
1.
First take your Moticam camera and remove the dust cap
from the bottom of the camera, being careful not to
introduce dust.
2.
Then take the MA15604 C-mount lens and attach it to the
Moticam camera. Then set to the side.
3.
Remove the top camera port cover of the microscope, by
loosening the three 1.5mm hex screws.
1.5mm hex
wrench purchased separately.
4.
Take the Moticam with attached C-Mount and slide down
into the open top camera port.
5.
Tighten the three 1.5mm hex screws back around the C-
mount to secure it in place.
6.
Then follow the camera instructions.