5 Specification
pH
s
imulation
v
alue
, normalised to 25°C
1
.00 pH, 1.68 pH, 4.01 pH, 6.86 pH, 7.00 pH, 9.18 pH, 10.01
pH, 12.45 pH
pH Simulation Accuracy (with supplied
cable)
+/- 0.02 pH
mV Simulation value
1800 mV, 900 mV, 390 mV, -390 mV, -900 mV, -1800 mV
mV Simulation Accuracy (with supplied
cable)
+/- 1 mV
High Impedance Test (Hi Z)
1 G ohm at 4.01 pH & 10.01 pH
Buffer Set Standard Simulation Value
USA, NIST
Display
LED, Orange lighting
Output
BNC Connector
Power Requirements
4 ‘AAA’ 1.5V Batteries
Operating Temperature
0 to 40°C
Relative Humidity
80% up to 31°C decreasing linearly to 50% at 40°C
Dimensions
14 X 17 X 3.5 cm
Weight
200g (Boxed)
6 Accessories
Item
Instruments Order Code
pH Simulator with 1m BNC to BNC Cable M2740
1m BNC to 13mm pH-plug Cable
M2740 13mm ph-plug cable