M2118
Operating Manual
V1.06 © 17.10.2011 MOSTEC AG, CH-4410 Liestal
5
E. Adjusting a tare value
Note: The measuring range is reduced by the set tare!
1. Adjusting a tare value with the programming buttons
1. Press and hold the button <ENTER> and <+>.
ð
The display shows
tarE
and the actual measuring value, after 3 seconds tare is adjusted.
ð
The display shows the actual measuring value.
2. Switch off tare with the programming buttons
1. Press and hold the button <ENTER> and <+>.
ð
The display shows
tarE
and the actual measuring value, after 3 seconds tare is switched off.
ð
The display shows the actual measuring value.
3. Adjusting a tare value with the menu
1. Press and hold the button <ENTER> and <ESC> for 4 seconds..
ð
The display shows
SP_1
indicating the user menu.
2. Move to menu item TARE. To move, use the button <+> or <->, then press <ENTER>.
3. Switch on or switch off tare. To move, use the button <+> or <->, then press <ENTER>.
4. Press the button <ESC>
ð
the display shows the actual measuring value.
F. Adjusting the output signal range with programming software
C
onnect the link cable with the Computer and the display. (see I, page 6)
1. start "M
pro
"
2. Choose "M2118" in menu "Instrument" <A>
3. Choose "Linear conversion table" under "Range" <B>
4. Choose "Adjust the output signal" <D>
5. Enter the values for the limit contacts and all other settings (hysteresis etc.)
6. Choose "Program the limit contacts"
<A>
<B>
<C>
<D>